Comparison of Postoperative Refractive Outcomes: IOLMaster ® Versus Immersion Ultrasound
Authors:
- Byung-Ju JungBJByung-Ju Jung
- Tae-Hoon OhTOTae-Hoon Oh
- Yong-Soo ByunYBYong-Soo Byun
- Choun-Ki JooCJChoun-Ki Joo
- 1 November 2012
- journal article
- Published by SLACK, Inc. in Ophthalmic Surgery, Lasers and Imaging Retina
- Vol. 43 (6) , 496-499
Abstract
BACKGROUND AND OBJECTIVE: To compare the postoperative refractive outcomes between IOLMaster biometry (Carl Zeiss Meditec, Inc., Dublin, CA) and immersion ultrasound biometry for axial length measurements. PATIENTS AND METHODS: Refractive outcomes in 354 eyes were compared using the IOLMaster and the immersion ultrasound biometry. Predicted refraction was determined using manual keratometry and the SRK-T formula with personalized A-constant. RESULTS: The axial lengths measured using the IOLMaster and immersion ultrasound were 24.49 ± 2.11 and 24.46 ± 2.11 mm, respectively, and the difference was significant ( P < .05). The mean errors were 0.000 ± 0.578 D with the IOLMaster, and 0.000 ± 0.599 D with the immersion ultrasound, but the difference was not significant. The mean absolute error was smaller with the IOLMaster than with immersion ultrasound (0.463 ± 0.341 vs 0.479 ± 0.359 D), but the difference was not significant. CONCLUSION: IOLMaster biometry yields highly accurate results in cataract surgery. However, if the IOLMaster is unavailable, immersion ultrasound biometry with personalized intraocular lens constants is an acceptable alternative.Keywords
This publication has 21 references indexed in Scilit:
- Comparison of refractive outcomes using immersion ultrasound biometry and IOLMaster biometryClinical & Experimental Ophthalmology, 2009
- Accuracy of modern intraocular lens power calculation formulas in refractive lens exchange for high myopia and high hyperopiaJournal of Cataract & Refractive Surgery, 2009
- Biometry and intraocular lens power calculationCurrent Opinion in Ophthalmology, 2008
- Partial coherence laser interferometry vs conventional ultrasound biometry in intraocular lens power calculationsEye, 2002
- A new non-contact optical device for ocular biometryBritish Journal of Ophthalmology, 2002
- Accuracy and reproducibility of biometry using partial coherence interferometryJournal of Cataract & Refractive Surgery, 2002
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Contact or immersion technique for axial length measurement?Australian and New Zealand Journal of Ophthalmology, 1999
- Refractive aspects of cataract surgeryCurrent Opinion in Ophthalmology, 1998
- Clear Corneal Cataract Surgery and the Correction of Myopia, Hyperopia, and AstigmatismOphthalmology, 1997