A simple, low‐cost and robust capillary zone electrophoresis method with capacitively coupled contactless conductivity detection for the routine determination of four selected penicillins in money‐constrained laboratories
- 12 April 2018
- journal article
- research article
- Published by Wiley in Electrophoresis
- Vol. 39 (20) , 2521-2529
Abstract
A simple and robust capillary zone electrophoresis method was developed and validated for the determination of amoxicillin and clavulanate, ampicillin, phenoxymethyl penicillin (Pen V) as well as flucloxacillin. Capacitively coupled contactless conductivity detection was employed as detection mode that makes CE a simple and economic tool for money‐constrained laboratories. The developed method is straightforward and user‐friendly. It offers good sensitivity and sufficient selectivity for the routine assay of the selected penicillins. The repeatabilities were <1.9% RSD for relative peak areas and 0.995) within the 80–120% range of the target concentration (0.5 mg/mL) for each antibiotic. The accuracy of the method, evaluated by standard fortification at three levels, was good for all the analytes. An extended robustness study was performed by varying ±10% of the optimum value of TRIS concentration, l‐histidine concentration and temperature in a full factorial design at two levels. This was to evaluate larger than usual variability of factors on the assay value, in order to better cover potential global variance in lab conditions and equipment. Finally, the method was applied for the determination of percent (%) content of all antibiotics in available formulations.Keywords
Funding Information
- KU Leuven
This publication has 24 references indexed in Scilit:
- Portable Capillary Electrophoresis Instrument with Automated Injector and Contactless Conductivity DetectionAnalytical Chemistry, 2013
- Capillary electrophoresis and isoelectric focusing in peptide and protein analysisProteomics, 2012
- Reliable low-cost capillary electrophoresis device for drug quality control and counterfeit medicinesJournal of Pharmaceutical and Biomedical Analysis, 2010
- Applications of capillary electrophoresis with capacitively coupled contactless conductivity detection (CE‐C4D) in pharmaceutical and biological analysisBiomedical Chromatography, 2010
- Capillary electrophoresis with contactless conductivity detection coupled to a sequential injection analysis manifold for extended automated monitoring applicationsAnalytica Chimica Acta, 2010
- Fundamentals of electrochemical detection techniques for CE and MCEElectrophoresis, 2009
- Determination of pharmaceuticals classified as emerging pollutants using capillary electrophoresis with capacitively coupled contactless conductivity detectionElectrophoresis, 2008
- Understanding Capacitively Coupled Contactless Conductivity Detection in Capillary and Microchip Electrophoresis. Part 1. FundamentalsElectroanalysis, 2005
- Fundamental aspects of contactless conductivity detection for capillary electrophoresis. Part I: Frequency behavior and cell geometryElectrophoresis, 2004
- Fundamental aspects of contactless conductivity detection for capillary electrophoresis. Part II: Signal‐to‐noise ratio and stray capacitanceElectrophoresis, 2004