Pulse transit time measurement using seismocardiogram and in-ear acoustic sensor
Authors:
- 1 October 2016
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 188-191
Abstract
This paper presents a new setup and method of measuring pulse transit time (PTT) using seismocardiographic data from a MEMS accelerometer and acoustic recordings from a commercial microphone placed inside the ear. A DSP system is designed and implemented in MATLAB, and experimentally tested on 5 healthy adult subjects at rest. Failure conditions are also examined to evaluate the robustness of the system. Our results suggest that the proposed technique could be a highly-relevant method for PTT measurements, which could be further translated into blood pressure (BP) values.Keywords
This publication has 13 references indexed in Scilit:
- An in-ear pulse wave velocity measurement system using heart sounds as time referenceCurrent Directions in Biomedical Engineering, 2015
- Toward Ubiquitous Blood Pressure Monitoring via Pulse Transit Time: Theory and PracticeIEEE Transactions on Biomedical Engineering, 2015
- From Wearable Sensors to Smart Implants-–Toward Pervasive and Personalized HealthcareIEEE Transactions on Biomedical Engineering, 2015
- Unobtrusive Sensing and Wearable Devices for Health InformaticsIEEE Transactions on Biomedical Engineering, 2014
- Attenuation of Systolic Blood Pressure and Pulse Transit Time Hysteresis During Exercise and Recovery in Cardiovascular PatientsIEEE Transactions on Biomedical Engineering, 2013
- Comparison of bilateral pulse arrival time before and after induced vasodilation by axillary blockPhysiological Measurement, 2012
- A wearable vital signs monitor at the ear for continuous heart rate and Pulse Transit Time measurementsAnnual International Conference of the IEEE Engineering in Medicine and Biology Society, 2012
- Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressureJournal of Applied Physiology, 2011
- Pulse transit time measured from the ECG: an unreliable marker of beat-to-beat blood pressureJournal of Applied Physiology, 2006
- Variability of photoplethysmographyperipheral pulse measurementsat the ears, thumbs and toesIEE Proceedings - Science, Measurement and Technology, 2000