Utilization of Assay Performance Characteristics to Estimate Hemoglobin A1c Result Reliability
Authors:
- 1 August 2014
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 60 (8) , 1073-1079
Abstract
BACKGROUND Allowable total error (TEa) goals for hemoglobin (Hb) A1c require minimal assay imprecision and bias and implementation of a robust QC monitoring program. Here, we compare the combined influence on the risk of reporting unreliable results of TEa goals, a routine QC practice, and assay performance characteristics of 6 Hb A1c instruments across 4 academic medical centers. METHODS The CLSI protocols EP-5 and EP-9 were applied to investigate Hb A1c result imprecision and bias on the Variant II Turbo and Variant II (Bio-Rad), G8 (Tosoh), Capillarys 2 Flex Piercing (Sebia), COBAS Integra 800 (Roche), and DCA Vantage (Siemens). Patient-weighted σ values and the risk of reporting unreliable Hb A1c results were determined for each assay at TEa specifications of 5%, 6%, and 7%. RESULTS A large range of patient-weighted σ values spanning 0.5 orders of magnitude at a 6% TEa was observed. Although imprecision for all instruments was <3%, bias impacted the majority of the σ changes observed. Estimates for reporting unreliable results varied almost 500-fold based on analytical performance alone. CONCLUSIONS Considerable differences in the probability of reporting unreliable Hb A1c results between different NGSP (formerly the National Glycohemoglobin Standardization Program)-certified platforms were observed. At a 6% TEa, our study indicates all but the Capillarys 2 Flex Piercing requires that the maximum affordable QC be run. Risk estimates for individual laboratories' Hb A1c methods can be used to assess QC practices and residual risk of an unreliable Hb A1c result.Keywords
This publication has 7 references indexed in Scilit:
- Perspectives on Quality Control, Risk Management, and Analytical Quality ManagementClinics In Laboratory Medicine, 2013
- Standards of Medical Care in Diabetes—2013Diabetes Care, 2012
- Guidelines and Recommendations for Laboratory Analysis in the Diagnosis and Management of Diabetes MellitusClinical Chemistry, 2011
- Status of Hemoglobin A1c Measurement and Goals for Improvement: From Chaos to Order for Improving Diabetes CareClinical Chemistry, 2011
- Practical application of Sigma Metrics QC procedures in clinical chemistryClinical Chemistry and Laboratory Medicine (CCLM), 2011
- Six of Eight Hemoglobin A1c Point-of-Care Instruments Do Not Meet the General Accepted Analytical Performance CriteriaClinical Chemistry, 2010
- Assessing the Impact of the Frequency of Quality Control Testing on the Quality of Reported Patient ResultsClinical Chemistry, 2008