Single scan parameterization of space-variant point spread functions in image space via a printed array: the impact for two PET/CT scanners
Authors:
- F A KotasidisFKF A Kotasidis
- J C MatthewsJMJ C Matthews
- G I AngelisGAG I Angelis
- P J NoonanPNP J Noonan
- A JacksonAJA Jackson
- P PricePPP Price
- W R LionheartWLW R Lionheart
- A J ReaderARA J Reader
- 13 April 2011
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 56 (10) , 2917-2942
Abstract
Incorporation of a resolution model during statistical image reconstruction often produces images of improved resolution and signal-to-noise ratio. A novel and practical methodology to rapidly and accurately determine the overall emission and detection blurring component of the system matrix using a printed point source array within a custom-made Perspex phantom is presented. The array was scanned at different positions and orientations within the field of view (FOV) to examine the feasibility of extrapolating the measured point source blurring to other locations in the FOV and the robustness of measurements from a single point source array scan. We measured the spatially-variant image-based blurring on two PET/CT scanners, the B-Hi-Rez and the TruePoint TrueV. These measured spatially-variant kernels and the spatially-invariant kernel at the FOV centre were then incorporated within an ordinary Poisson ordered subset expectation maximization (OP-OSEM) algorithm and compared to the manufacturer's implementation using projection space resolution modelling (RM). Comparisons were based on a point source array, the NEMA IEC image quality phantom, the Cologne resolution phantom and two clinical studies (carbon-11 labelled anti-sense oligonucleotide [(11)C]-ASO and fluorine-18 labelled fluoro-l-thymidine [(18)F]-FLT). Robust and accurate measurements of spatially-variant image blurring were successfully obtained from a single scan. Spatially-variant resolution modelling resulted in notable resolution improvements away from the centre of the FOV. Comparison between spatially-variant image-space methods and the projection-space approach (the first such report, using a range of studies) demonstrated very similar performance with our image-based implementation producing slightly better contrast recovery (CR) for the same level of image roughness (IR). These results demonstrate that image-based resolution modelling within reconstruction is a valid alternative to projection-based modelling, and that, when using the proposed practical methodology, the necessary resolution measurements can be obtained from a single scan. This approach avoids the relatively time-consuming and involved procedures previously proposed in the literature.Keywords
This publication has 24 references indexed in Scilit:
- Non-Gaussian space-variant resolution modelling for list-mode reconstructionPhysics in Medicine & Biology, 2010
- Image-based point spread function implementation in a fully 3D OSEM reconstruction algorithm for PETPhysics in Medicine & Biology, 2010
- Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluationPhysics in Medicine & Biology, 2010
- Iterative image reconstruction for positron emission tomography based on a detector response function estimated from point source measurementsPhysics in Medicine & Biology, 2009
- Analytic system matrix resolution modeling in PET: an application to Rb-82 cardiac imagingPhysics in Medicine & Biology, 2008
- Impact of Image-Space Resolution Modeling for Studies with the High-Resolution Research TomographJournal of Nuclear Medicine, 2008
- Production of radioactive quality assurance phantoms using a standard inkjet printerPhysics in Medicine & Biology, 2007
- Object Dependency of Resolution in Reconstruction Algorithms With Interiteration Filtering Applied to PET DataIEEE Transactions on Medical Imaging, 2004
- EM algorithm system modeling by image-space techniques for PET reconstructionIEEE Transactions on Nuclear Science, 2003
- One-pass list-mode EM algorithm for high-resolution 3-D PET image reconstruction into large arraysIEEE Transactions on Nuclear Science, 2002