Validation of SPAMM Tagged MRI Based Measurement of 3D Soft Tissue Deformation
Authors:
Preprint
Abstract
This study presents and validates a novel (non-ECG-triggered) MRI sequence based on SPAtial Modulation of the Magnetization (SPAMM) to non-invasively measure 3D (quasi-static) soft tissue deformations using only six acquisitions (three static and three indentations). In current SPAMM tagged MRI approaches data is typically constructed from many repeated motion cycles. This has so far restricted its application to the measurement of highly repeatable and periodic movements (e.g. cardiac deformation). In biomechanical applications where soft tissue deformation is artificially induced, often by indentation, significant repeatability constraints exist and, for clinical applications, discomfort and health issues generally preclude a large number of repetitions.Keywords
All Related Versions
- Version 1, 2016-04-22, ArXiv
- Published version: Medical Physics, 38 (3), 1248.
This publication has 41 references indexed in Scilit:
- 3D left ventricular strain from unwrapped harmonic phase measurementsJournal of Magnetic Resonance Imaging, 2010
- Characterization of engineered tissue construct mechanical function by magnetic resonance imagingJournal of Tissue Engineering and Regenerative Medicine, 2009
- MRI of myocardial infarction with tissue taggingCurrent Cardiovascular Imaging Reports, 2009
- MR elastography of breast lesions: Understanding the solid/liquid duality can improve the specificity of contrast‐enhanced MR mammographyMagnetic Resonance in Medicine, 2007
- Noninvasive Determination of Ligament Strain with Deformable Image RegistrationAnnals of Biomedical Engineering, 2007
- T1, T2 relaxation and magnetization transfer in tissue at 3TMagnetic Resonance in Medicine, 2005
- Tracking MR tag surfaces using a spatiotemporal filter and interpolatorInternational Journal of Imaging Systems and Technology, 1999
- DENSE: Displacement Encoding with Stimulated Echoes in Cardiac Functional MRIJournal of Magnetic Resonance, 1999
- MRI of myocardial function: motion tracking techniquesMagnetic Resonance Imaging, 1996
- A flow velocity zeugmatographic interlace for NMR imaging in humansMagnetic Resonance Imaging, 1982