Accuracy of model-based tracking of knee kinematics and cartilage contact measured by dynamic volumetric MRI
- 4 July 2016
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 38 (10) , 1131-1135
Abstract
The purpose of this study was to determine the accuracy of knee kinematics and cartilage contact measured by volumetric dynamic MRI. A motor-actuated phantom drove femoral and tibial bone segments through cyclic 3D motion patterns. Volumetric images were continuously acquired using a 3D radially undersampled cine spoiled gradient echo sequence (SPGR-VIPR). Image data was binned based on position measured via a MRI-compatible rotary encoder. High-resolution static images were segmented to create bone models. Model-based tracking was performed by optimally registering the bone models to the volumetric images at each frame of the SPGR-VIPR series. 3D tibiofemoral translations and orientations were reconstructed, and compared to kinematics obtained by tracking fiducial markers. Imaging was repeated on a healthy subject who performed cyclic knee flexion-extension. Cartilage contact for the subject was assessed by measuring the overlap between articular cartilage surfaces. Model-based tracking was able to track tibiofemoral angles and translations with precisions less than 0.8° and 0.5mm. These precisions resulted in an uncertainty of less than 0.5mm in cartilage contact location. Dynamic SPGR-VIPR imaging can accurately assess in vivo knee kinematics and cartilage contact during voluntary knee motion performed in a MRI scanner. This technology could facilitate the quantitative investigation of links between joint mechanics and the development of osteoarthritis.Keywords
Funding Information
- NIH (EB015410)
- NIH (AR062733)
This publication has 36 references indexed in Scilit:
- Do dynamic‐based MR knee kinematics methods produce the same results as static methods?Magnetic Resonance in Medicine, 2012
- Cartilage in Anterior Cruciate Ligament–Reconstructed Knees: MR Imaging T1ρand T2—Initial Experience with 1-year Follow-upRadiology, 2011
- Measurement of in vivo anterior cruciate ligament strain during dynamic jump landingJournal of Biomechanics, 2011
- Assessing the accuracy and precision of musculoskeletal motion tracking using cine-PC MRI on a 3.0T platformJournal of Biomechanics, 2011
- Differences in tibial rotation during walking in ACL reconstructed and healthy contralateral kneesJournal of Biomechanics, 2010
- Evaluation of Kinematics of Anterior Cruciate Ligament-Deficient Knees with Use of Advanced Imaging Techniques, Three-Dimensional Modeling Techniques, and RoboticsJournal of Bone and Joint Surgery, 2009
- Validation of three-dimensional model-based tibio-femoral tracking during runningMedical Engineering & Physics, 2008
- Magnetic resonance imaging (MRI) of articular cartilage in knee osteoarthritis (OA): morphological assessmentOsteoarthritis and Cartilage, 2006
- Composition of MRI phantom equivalent to human tissuesMedical Physics, 2005
- Determining the movements of the skeleton using well-configured markersJournal of Biomechanics, 1993