Monte Carlo simulations to evaluate geometric variations of a new Iodine-125 seed
- 18 February 2026
- journal article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 12 (2) , 025010
Abstract
The Nuclear and Energy Research Institute and the National Nuclear Energy Commission (IPEN-CNEN/SP) are developing a new Iodine-125 seed for use in brachytherapy. The manufacturing process includes cutting the titanium tube, welding the ends, and performing leak tests. Seed approval is based on a microscopic analysis of its external geometry. The dosimetric methodology follows Task Group 43 (TG-43), which adopts the Monte Carlo method as the standard simulation technique. However, TG-43 does not account for geometric variations that may occur during the manufacturing process. This study investigated two types of variation: (1) the seed width after welding, and (2) the random positions that the core may assume within the seed. A total of 100 simulations were performed for each variation. The width was measured experimentally using a caliper, after welding, while the core position was modeled using a random number generator, since it is not feasible to measure it directly. These values were used to generate 201 input files for simulations in MCNP 6.2: 100 for each investigated parameter and one for the reference geometry (IPEN-CNEN/SP seed). The results were analyzed using MATLAB through bidimensional matrices of 101 × 101 points. For each point in the matrix, the mean, standard deviation, relative difference (compared to the reference geometry), and Type A uncertainty were calculated. The main results were: a standard deviation of 0.6% and a relative difference of 11% for the seed width; and a standard deviation of 50% and a relative difference of 25% for the core position. The Type A uncertainty was below 0.3% in all cases.Keywords
Funding Information
- Conselho Nacional de Desenvolvimento Científico e Tecnológico (406761/2022-1)
- IPEN/CNEN (2020.06, 37)
- Centros de Pesquisa, Inovação e Difusão, Fundação Amazônia Paraense de Amparo à Pesquisa (50332-0 and 2020/07065-4.)
This publication has 9 references indexed in Scilit:
- Monte Carlo simulation to assess free space and end-weld thickness variation effects on dose rate for a new Ir-192 brachytherapy sourceApplied Radiation and Isotopes, 2021
- Dose-rate constant and air-kerma strength evaluation of a new 125I brachytherapy source using Monte CarloBrazilian Journal of Radiation Sciences, 2021
- RECORDS: improved Reporting of montE CarlO RaDiation transport Studies: Report of the AAPM Research Committee Task Group 268Medical Physics, 2017
- Report of the Task Group 186 on model‐based dose calculation methods in brachytherapy beyond the TG‐43 formalism: Current status and recommendations for clinical implementationMedical Physics, 2012
- Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculationsMedical Physics, 2004
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Effects of gold and silver backings on the dose rate around an 125I seedMedical Physics, 1990
- Monte Carlo evaluation of specific dose constants in water for 125I seedsMedical Physics, 1988
- A Monte Carlo derivation of parameters for use in the tissue dosimetry of medium and low energy nuclidesBritish Journal of Radiology, 1982