Class E driver for transcutaneous power and data link for implanted electronic devices
Authors:
- 1 January 1992
- journal article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 30 (1) , 69-75
Abstract
Magnetic transcutaneous coupling is frequently used for power and data transfer to implanted electronic devices. The paper describes a transmitter/coil driver based on the class E topology. The development of a ‘high-Q approximation’ simplifies the design procedure. A method of data modulation using synchronous frequency shifting is decribed. The class E circuit shows great promise, especially for circuits with unusually low coefficients of coupling. Transmitter coil currents of several amperes, at radio frequencies, with relatively low active device power dissipation, can easily be obtained.Keywords
This publication has 8 references indexed in Scilit:
- RF powering of millimeter- and submillimeter-sized neural prosthetic implantsIEEE Transactions on Biomedical Engineering, 1988
- MINIMOS 3: A MOSFET simulator that includes energy balanceIEEE Transactions on Electron Devices, 1987
- Radio-Frequency Coils in Implantable Devices: Misalignment Analysis and Design ProcedureIEEE Transactions on Biomedical Engineering, 1987
- A Wide-Band Efficient Inductive Transdennal Power and Data Link with Coupling Insensitive GainIEEE Transactions on Biomedical Engineering, 1987
- Interface states of modulation-doped AlGaAs/GaAs heterostructuresIEEE Transactions on Electron Devices, 1987
- Analysis of resonant coupled coils in the design of radio frequency transcutaneous linksMedical & Biological Engineering & Computing, 1983
- The maximum zero-voltage tunnel current and its fluctuations for lead—lead oxide—lead Josephson tunnel junctionsIEEE Transactions on Electron Devices, 1980
- Class E-A new class of high-efficiency tuned single-ended switching power amplifiersIEEE Journal of Solid-State Circuits, 1975