Monte Carlo-based diode design for correction-less small field dosimetry
Authors:
- P H CharlesPCP H Charles
- S B CroweSCS B Crowe
- T KairnTKT Kairn
- R T KnightRKR T Knight
- B HillBHB Hill
- J KennyJKJ Kenny
- C M LangtonCLC M Langton
- J V TrappJTJ V Trapp
- 13 June 2013
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 58 (13) , 4501-4512
Abstract
Due to their small collecting volume, diodes are commonly used in small field dosimetry. However, the relative sensitivity of a diode increases with decreasing small field size. Conversely, small air gaps have been shown to cause a significant decrease in the sensitivity of a detector as the field size is decreased. Therefore, this study uses Monte Carlo simulations to look at introducing air upstream to diodes such that they measure with a constant sensitivity across all field sizes in small field dosimetry. Varying thicknesses of air were introduced onto the upstream end of two commercial diodes (PTW 60016 photon diode and PTW 60017 electron diode), as well as a theoretical unenclosed silicon chip using field sizes as small as 5 mm × 5 mm. The metric D(w,Q)/D(Det,Q) used in this study represents the ratio of the dose to a point of water to the dose to the diode active volume, for a particular field size and location. The optimal thickness of air required to provide a constant sensitivity across all small field sizes was found by plotting D(w,Q)/D(Det,Q) as a function of introduced air gap size for various field sizes, and finding the intersection point of these plots. That is, the point at which D(w,Q)/D(Det,Q) was constant for all field sizes was found. The optimal thickness of air was calculated to be 3.3, 1.15 and 0.10 mm for the photon diode, electron diode and unenclosed silicon chip, respectively. The variation in these results was due to the different design of each detector. When calculated with the new diode design incorporating the upstream air gap, k(f(clin),f(msr))(Q(clin),Q(msr)) was equal to unity to within statistical uncertainty (0.5%) for all three diodes. Cross-axis profile measurements were also improved with the new detector design. The upstream air gap could be implanted on the commercial diodes via a cap consisting of the air cavity surrounded by water equivalent material. The results for the unclosed silicon chip show that an ideal small field dosimetry diode could be created by using a silicon chip with a small amount of air above it.Keywords
This publication has 14 references indexed in Scilit:
- The effect of very small air gaps on small field dosimetryPhysics in Medicine & Biology, 2012
- Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterizationPhysics in Medicine & Biology, 2012
- Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife systemPhysics in Medicine & Biology, 2012
- Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectorsMedical Physics, 2011
- Calculation of for several small detectors and for two linear accelerators using Monte Carlo simulationsMedical Physics, 2011
- Experimental small field 6MV output ratio analysis for various diode detector and accelerator combinationsRadiotherapy and Oncology, 2011
- Adapting a generic BEAMnrc model of the BrainLAB m3 micro-multileaf collimator to simulate a local collimation devicePhysics in Medicine & Biology, 2010
- Technical Note: Modeling a complex micro-multileaf collimator using the standard BEAMnrc distributionMedical Physics, 2010
- The influence of small field sizes, penumbra, spot size and measurement depth on perturbation factors for microionization chambersPhysics in Medicine & Biology, 2009
- A new formalism for reference dosimetry of small and nonstandard fieldsMedical Physics, 2008