Use of the FBX dosimeter for the quality assurance of60Co and high energy teletherapy machines by mail
Authors:
- 1 November 1992
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 37 (11) , 2095-2102
Abstract
The concentrations of the solutes in the standard FBX dosimeter were changed in order to reduce the rate of thermal oxidation of ferrous ions in the solution. The modified solution contained 10-4 mol dm-3 ferrous ions. 10-3 mol dm-3 benzoic acid and 5*10-5 mol dm-3 xylenol orange in 0.04 mol dm-3 aqueous aerated sulphuric acid. The rates of thermal oxidation of ferrous ions do not differ significantly when the solution is stored either in glass or polypropylene containers. The error due to the post-irradiation delay in dose estimation depends on the dose and a dose of about 5 Gy is most suitable. When the solution stored up to about one month before irradiation and about one month after irradiation, the error in dose estimation is less than +or-2%. The blank and the irradiated solutions should be stored in similar containers.Keywords
This publication has 7 references indexed in Scilit:
- G(Fe3+) values in the FBX dosemeterPhysics in Medicine & Biology, 1985
- Acid and xylenol orange effects in the FBX dosimeterThe International Journal of Applied Radiation and Isotopes, 1983
- Use of the FBX dosemeter for the calibration of cobalt-60 and high-energy teletherapy machinesPhysics in Medicine & Biology, 1982
- Dose-rate and fractionation studies with FBX dosimeterThe International Journal of Applied Radiation and Isotopes, 1981
- Radiation Chemistry of the Ferrous Sulfate-Benzoic Acid-Xylenol Orange SystemRadiation Research, 1978
- Consistency of ferrous sulphate-benzoic acid-xylenol orange dosimeterThe International Journal of Applied Radiation and Isotopes, 1974
- Reconsideration of resolution in EM autoradiography using a biological line sourceJournal of Microscopy, 1973