Does image quality matter? Impact of resolution and noise on mammographic task performance
Authors:
- Jay A. BakerJBJay A. Baker
- David M. DelongDDDavid M. Delong
- Jeff P. JohnsonJJJeff P. Johnson
- Ehsan SameiESEhsan Samei
- 24 September 2007
- journal article
- Published by Wiley in Medical Physics
- Vol. 34 (10) , 3971-3981
Abstract
The purpose of this study was to examine the effects of different resolution and noise levels on task performance in digital mammography. This study created an image set with images at three different resolution levels, corresponding to three digital display devices, and three different noise levels, with noise magnitudes similar to full clinical dose, half clinical dose, and quarter clinical dose. The images were read by five experienced breast imaging radiologists. The data were then analyzed to compute two accuracy statistics (overall classification accuracy and lesion detection accuracy) and performance at four diagnostic tasks (detection of microcalcifications, benign masses, malignant masses, and discrimination of benign and malignant masses). Human observer results showed decreasing display resolution had little effect on overall classification accuracy and individual diagnostic task performance, but increasing noise caused overall classification accuracy to decrease by a statistically significant 21% as the breast dose went to one quarter of its normal clinical value. The noise effects were most prominent for the tasks of microcalcification detection and mass discrimination. When the noise changed from full clinical dose to quarter clinical dose, the microcalcification detection performance fell from 89% to 67% and the mass discrimination performance decreased from 93% to 79%, while malignant mass detection performance remained relatively constant with values of 88% and 84%, respectively. As a secondary aim, the image set was also analyzed by two observer models to examine whether their performance was similar to humans. Observer models differed from human observers and each other in their sensitivity to resolution degradation and noise. The primary conclusions of this study suggest that quantum noise appears to be the dominant image quality factor in digital mammography, affecting radiologist performance much more profoundly than display resolution.Keywords
Funding Information
- National Institutes of Health (NIH R21‐CA95308)
- Medical Research and Materiel Command (USAMRMC W81XWH‐04‐1‐0323)
This publication has 60 references indexed in Scilit:
- Identification of simulated microcalcifications in white noise and mammographic backgroundsMedical Physics, 2006
- Simulation of Mammographic LesionsAcademic Radiology, 2006
- Image quality performance of liquid crystal display systems: Influence of display resolution, magnification and window settings on contrast-detail detectionEuropean Journal of Radiology, 2006
- On-Axis and Off-Axis Viewing of Images on CRT Displays and LCDsAcademic Radiology, 2005
- Imaging performance of amorphous selenium based flat-panel detectors for digital mammography: Characterization of a small area prototype detectorMedical Physics, 2003
- Determining the Area under the ROC Curve for a Binary Diagnostic TestMedical Decision Making, 2000
- REGRET GRAPHS, DIAGNOSTIC UNCERTAINTY AND YOUDEN'S INDEXStatistics in Medicine, 1996
- Statistically defined backgrounds:performance of a modified nonprewhitening observer modelJournal of the Optical Society of America A, 1994
- The Efficacy of Diagnostic ImagingMedical Decision Making, 1991
- ROC Methodology in Radiologic ImagingInvestigative Radiology, 1986