Stray Electrical Currents in Laparoscopic Instruments Used in da Vinci ® Robot-Assisted Surgery: An In Vitro Study
- 1 September 2011
- journal article
- research article
- Published by SAGE Publications in Journal of Endourology
- Vol. 25 (9) , 1513-1517
Abstract
Background and Purpose: The da Vinci® surgical system requires the use of electrosurgical instruments. The re-use of such instruments creates the potential for stray electrical currents from capacitive coupling and/or insulation failure. We used objective measures to report the prevalence and magnitude of such stray currents. Materials and Methods: Thirty-seven robotic instruments were tested using an electrosurgical unit (ESU) at pure coagulation and cut waveforms at four different settings. Conductive gel-coated instruments were tested at 40W, 80W, and maximum ESU output (coagulation 120W, cut 300W). The magnitude of stray currents was measured by an electrosurgical analyzer. Results: At coagulation waveform in open air, 86% of instruments leaked a mean of 0.4W. In the presence of gel-coated instruments, stray currents were detected in all instruments with means (and standard deviation) of 3.4W (±2), 4.1W (±2.3), and 4.1W (±2.3) at 40W, 80W, and 120W, respectively. At cut waveform in open air, none of the instruments leaked current, while gel-coated instruments leaked a mean of 2.2W (±1.3), 2.2W (±1.9) and 3.2W (±1.9) at 40W, 80W, and 300W, respectively. Conclusions: All tested instruments in our study demonstrated energy leakage. Stray currents were higher during coagulation (high voltage) waveforms, and the magnitude was not always proportionate to the ESU settings. Stray currents have the potential to cause electrical burns. We support the programmed end of life of da Vinci instruments on the basis of safety. Consideration should be given to alternate energy sources or the adoption of active electrode monitoring technology to all monopolar instruments.Keywords
This publication has 16 references indexed in Scilit:
- Training, Credentialing, Proctoring and Medicolegal Risks of Robotic Urological Surgery: Recommendations of the Society of Urologic Robotic SurgeonsJournal of Urology, 2009
- Insulation failure in laparoscopic instrumentsSurgical Endoscopy, 2009
- The Cost of Learning Robotic-Assisted ProstatectomyUrology, 2008
- Characterization and mitigation of stray radiofrequency currents during monopolar resectoscopic electrosurgeryJournal of Minimally Invasive Gynecology, 2006
- Mechanisms of thermal injury to the lower genital tract with radiofrequency resectoscopic surgeryJournal of Minimally Invasive Gynecology, 2006
- Delayed Perforation of the Large Bowel due to Thermal Injury during Laparoscopic CholecystectomyJournal of International Medical Research, 2005
- Laparoscopic and Robot Assisted Radical Prostatectomy: Establishment of a Structured Program and Preliminary Analysis of OutcomesJournal of Urology, 2002
- Electrophysical properties of electrosurgery and capacitive induced currentThe American Journal of Surgery, 2001
- Intended and stray radiofrequency electrical currents during resectoscopic surgeryThe Journal of the American Association of Gynecologic Laparoscopists, 2000
- The medicolegal risks of thermal injury during laparoscopic monopolar electrosurgeryJournal of Healthcare Risk Management, 1998