Abstract
The effects of the design of a radiographic system on the modulation transfer function (MTF) are studied with a specially developed computer program. The program simulates a digital radiographic system by using three parameters: sampling distance, sampling aperture, and the spread of the signal in the detector due to the interaction processes of the incoming photons. The signal spread is approximated by Gaussian distributions. The influence of the three parameters is studied on the presampling MTF and on the two extreme cases of the digital MTF: the maximum MTF and the minimum MTF. From theoretical data on the interaction processes, the resolution properties of an amorphous selenium flat-panel detector are simulated. The program is also used to simulate a measurement of the presampling MTF with the slit method, and the effect of the slit width on the measured presampling MTF is examined.