Design and calibration of a high-frequency oscillatory ventilator
Authors:
- W. MitznerWMW. Mitzner
- 1 January 1991
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 38 (2) , 214-218
Abstract
High-frequency ventilation (HFV) is a modality of mechanical ventilation which presents difficult technical demands to the clinical or laboratory investigator. The essential features of an ideal HFV system are described, including wide frequency range, control of tidal volume and mean airway pressure, minimal dead space, and high effective internal impedance. The design and performance of a high-frequency oscillatory ventilation system is described which approaches these requirements. The ventilator utilizes a linear motor regulated by a closed loop controller and driving a novel frictionless double-diaphragm piston pump. Finally, the ventilator performance is tested using the impedance model of Venegas [1].Keywords
This publication has 10 references indexed in Scilit:
- Gas mixing in lung model ventilated by high frequency oscillation: Effect of tidal volume, frequency and molecular diffusivityMedical & Biological Engineering & Computing, 1991
- Equivalent Circuit Analysis of High-Frequency Ventilators Including a New High-Impedance Flow-Interrupting VentilatorIEEE Transactions on Biomedical Engineering, 1986
- Resonant amplification of delivered volume during high-frequency ventilationJournal of Applied Physiology, 1986
- Measurement of peak systolic elastance in intact canine circulation with servo pumpAmerican Journal of Physiology-Heart and Circulatory Physiology, 1985
- Mean airway pressure and alveolar pressure during high-frequency ventilationJournal of Applied Physiology, 1984
- Physiological dead space during high-frequency ventilation in dogsJournal of Applied Physiology, 1984
- A new system for ventilating with high-frequency oscillationJournal of Applied Physiology, 1982
- A new ventilator for physiologic studies during high-frequency ventilationRespiration Physiology, 1982
- Tidal Volume and Frequency Dependence of Carbon Dioxide Elimination by High-Frequency VentilationNew England Journal of Medicine, 1981
- Effects of frequency, tidal volume, and lung volume on CO2 elimination in dogs by high frequency (2-30 Hz), low tidal volume ventilation.Journal of Clinical Investigation, 1981