Triple Resonance Methods to Improve Performance of Ion Trap Precursor and Neutral Loss Scans
- 13 April 2020
- journal article
- research article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 31 (5) , 1123-1131
Abstract
Two experiments are described that extend the capabilities of quadrupole ion trap mass spectrometers operated in the precursor and neutral loss scan mode. The first experiment, a triple resonance precursor ion scan is used to enhance sensitivity, selectivity, and molecular coverage. This method augments the ion trap precursor ion scan with application of a second excitation frequency to selectively activate first generation (MS2) product ions as they are formed and produce second generation (MS3) product ions which are then mass-selectively ejected with a third auxiliary signal and detected. This single mass analyzer experiment can be equated to performing the sequential precursor ion scan in a multiple analyzer system (Anal. Chem. 1990, 62, 1809-1818). The second capability demonstrated is ‘frequency tagging’, a method used to differentiate between ions ejected due to inherent instability under given trapping conditions – which causes artifacts during these scans - and ions that are resonantly ejected by the product ion ejection frequency. Beat frequencies are used to modulate resonance ejection peaks but conveniently do not modulate boundary ejection peaks. Frequency tagging provides a mechanism to identify the artifact peaks that are a consequence of operating at high trapping voltage (i.e. low mass cutoff) for optimal precursor/product ion selectivity. The experiment is demonstrated for precursor and for neutral loss scans.Keywords
Funding Information
- National Aeronautics and Space Administration (NNX16AJ25G)
- FLIR Systems, Inc.
This publication has 24 references indexed in Scilit:
- Single analyzer precursor scans using an ion trapRapid Communications in Mass Spectrometry, 2016
- Vacuum Ultraviolet Photodissociation and Fourier Transform–Ion Cyclotron Resonance (FT-ICR) Mass Spectrometry: RevisitedAnalytical Chemistry, 2016
- Ultraviolet photodissociation of protonated, fixed charge, and charge-reduced peptidesInternational Journal of Mass Spectrometry, 2015
- Characterization of the dehydration products due to thermal decomposition of peptides by liquid chromatography‐tandem mass spectrometryJournal of Mass Spectrometry, 2015
- Insight into the resonance ejection process during mass analysis through simulations for improved linear quadrupole ion trap mass spectrometer performanceInternational Journal of Mass Spectrometry, 2014
- High-Resolution Multistage MS, MS2, and MS3 Matrix-Assisted Laser Desorption/Ionization FT-ICR Mass Spectra of Peptides from a Single Laser ShotAnalytical Chemistry, 1996
- The non-linear ion trap. Part 5. Nature of non-linear resonances and resonant ion ejectionInternational Journal of Mass Spectrometry and Ion Processes, 1994
- The non-linear ion trap. Part 4. Mass selective instability scan with multipole superpositionInternational Journal of Mass Spectrometry and Ion Processes, 1993
- Systematic delineation of scan modes in multidimensional mass spectrometryAnalytical Chemistry, 1990
- Selected ion fragmentation with a tandem quadrupole mass spectrometerJournal of the American Chemical Society, 1978