Optimizing dual energy cone beam CT protocols for preclinical imaging and radiation research
- 1 January 2017
- journal article
- review article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 90 (1069) , 20160480
Abstract
The aim of this work was to investigate whether quantitative dual-energy CT (DECT) imaging is feasible for small animal irradiators with an integrated cone-beam CT (CBCT) system. The optimal imaging protocols were determined by analyzing different energy combinations and dose levels. The influence of beam hardening effects and the performance of a beam hardening correction (BHC) were investigated. In addition, two systems from different manufacturers were compared in terms of errors in the extracted effective atomic numbers (Zeff) and relative electron densities (ρe) for phantom inserts with known elemental compositions and relative electron densities. The optimal energy combination was determined to be 50 and 90 kVp. For this combination, Zeff and ρe can be extracted with a mean error of 0.11 and 0.010, respectively, at a dose level of 60 cGy. Quantitative DECT imaging is feasible for small animal irradiators with an integrated CBCT system. To obtain the best results, optimizing the imaging protocols is required. Well-separated X-ray spectra and a sufficient dose level should be used to minimize the error and noise for Zeff and ρe. When no BHC is applied in the image reconstruction, the size of the calibration phantom should match the size of the imaged object to limit the influence of beam hardening effects. No significant differences in Zeff and ρe errors are observed between the two systems from different manufacturers. This is the first study that investigates quantitative DECT imaging for small animal irradiators with an integrated CBCT system.Keywords
This publication has 26 references indexed in Scilit:
- Improved tissue assignment using dual-energy computed tomography in low-dose rate prostate brachytherapy for Monte Carlo dose calculationMedical Physics, 2016
- Dual energy CT in radiotherapy: Current applications and future outlookRadiotherapy and Oncology, 2016
- Dual-Energy Computed Tomography Imaging of Atherosclerotic Plaques in a Mouse Model Using a Liposomal-Iodine Nanoparticle Contrast AgentCirculation: Cardiovascular Imaging, 2013
- Simulation study on potential accuracy gains from dual energy CT tissue segmentation for low-energy brachytherapy Monte Carlo dose calculationsPhysics in Medicine & Biology, 2011
- Extracting atomic numbers and electron densities from a dual source dual energy CT scanner: Experiments and a simulation modelRadiotherapy and Oncology, 2011
- The importance of tissue segmentation for dose calculations for kilovoltage radiation therapyMedical Physics, 2011
- Sensitivity of low energy brachytherapy Monte Carlo dose calculations to uncertainties in human tissue compositionMedical Physics, 2010
- Guidelines for the welfare and use of animals in cancer researchBritish Journal of Cancer, 2010
- Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculationsPhysics in Medicine & Biology, 2008
- Tissue segmentation in Monte Carlo treatment planning: A simulation study using dual-energy CT imagesRadiotherapy and Oncology, 2008