Analytical evaluation of the novel Mindray high sensitivity cardiac troponin I immunoassay on CL-1200i
Authors:
- 8 January 2024
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 62 (7) , 1433-1437
Abstract
Objectives: The current study was designed to evaluate the analytical performance of the new Mindray highly sensitive cardiac troponin I (hs-cTnI) chemiluminescent immunoassay on Mindray CL-1200i, as a thorough validation of novel hs-cTnI methods is required before introduction into clinical practice. Methods: The evaluation of the analytical performance of this hs-cTnI immunoassay encompassed the calculation of the limit of blank (LOB), limit of detection (LOD), functional sensitivity, imprecision, linearity, 99th percentile upper reference limit (URL) and concordance with another previously validated hs-cTnI chemiluminescent immunoassay. Results: The LOB and LOD were 0.32 and 0.35 ng/L, whilst the functional sensitivity (expressed as cTnI value with <10 % imprecision), was 0.35 ng/L. The linearity was excellent throughout a wide range of clinically measurable values (r=1.00 between 0.8 and 9,726.9 ng/mL). The intra-assay, inter-assay and total imprecision were 1.1–1.3 %, 5.5–8.1 % and 5.6–8.2 %, respectively. The 99th percentile URL calculated using residual plasma from 246 ostensibly healthy blood donors was 9.2 ng/L (4.3 ng/L in women vs. 12.3 ng/L in men). The Spearman’s correlation between Mindray hs-cTnI and Access hs-TnI was 0.97, with mean bias of 7.2 % (95 % CI, 2.6–11.9 %). Conclusions: Although we failed to confirm the very optimistic analytical characteristics previously reported for this method, our evaluation of the novel Mindray hs-cTnI immunoassay on CL-1200i demonstrated that the overall performance is comparable to that of other commercially available hs-cTnI techniques, making it a viable alternative to other methods.Keywords
This publication has 10 references indexed in Scilit:
- Evaluation of the analytical and clinical performance of a new high-sensitivity cardiac troponin I assay: hs-cTnI (CLIA) assayClinical Chemistry and Laboratory Medicine (CCLM), 2023
- Rapid rule-in and rule-out protocols of acute myocardial infarction using hs-cTnI and hs-cTnT methodsClinical Chemistry and Laboratory Medicine (CCLM), 2023
- 2023 ESC Guidelines for the management of acute coronary syndromesEuropean Heart Journal, 2023
- Correction to: Heart Disease and Stroke Statistics—2023 Update: A Report From the American Heart AssociationCirculation, 2023
- High-sensitivity methods for cardiac troponins: The mission is not over yet2021
- Laboratory-related issues in the measurement of cardiac troponins with highly sensitive assaysClinical Chemistry and Laboratory Medicine (CCLM), 2020
- Analytical evaluation of the new Beckman Coulter Access high sensitivity cardiac troponin I immunoassayClinical Chemistry and Laboratory Medicine (CCLM), 2017
- Diagnostic algorithms for acute coronary syndrome—is one better than another?Annals of Translational Medicine, 2016
- Time Delay to Treatment and Mortality in Primary Angioplasty for Acute Myocardial InfarctionCirculation, 2004
- How to improve estimates of imprecision.Clinical Chemistry, 1984