Diagnostic Accuracy of Electrical Pulp Testing Compared to the Thermal in Primary Teeth

Authors:

Abstract
Aim: This study aimed to assess the diagnostic accuracy of electric pulp testing compared tothermal method in registering pulp status in primary molars.Subjects & Methods: Forty cooperative children with carious primary molars were includedin the study. The cold test was carried out using Ethyl Chloride on cotton pellet and was applied tothe middle third of the buccal surface until the child indicated that he had felt sensation. The heattest was carried out using heated gutta-percha. Before using the electric pulp tester (EPT) teethwere isolated and dried. The gold standard for pulp status was decided by direct inspection of thepresence or absence of bleeding after access cavity had been done. Data were statistically describedin terms of frequencies and percentages. Comparison between the different test results was doneusing McNemar test. Agreement was tested using kappa statistic. Accuracy was representedusing the terms sensitivity, specificity, +ve predictive value, -ve predictive value, and accuracy.Comparing accuracy was done using Yates corrected Chi squared test. p values less than 0.05 wasconsidered statistically significant.Results: The highest accuracy was found for EPT followed by heat and cold tests. The highesttrue negative response was recorded for EPT (47.5%), then heat and cold (35%) and (30%)respectively. Sensitivity for EPT, heat and cold was 84.2%, 94.74% and 84.21 respectively while,specificity was 90.48%, 66.67% and 57.14%.Conclusions: EPT can be used as a sensible test for diagnosing pulp status in primary teeth.