Repeatability of Pattern Electroretinogram Measurements Using a New Paradigm Optimized for Glaucoma Detection
- 1 August 2009
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Journal of Glaucoma
- Vol. 18 (6) , 437-442
Abstract
Purpose To determine the within-trial and between-trial repeatability of pattern electroretinogram (PERG) measurements in healthy and patient eyes, using a new clinical instrument, the PERGLA. Study Design In all, 70 eyes of 35 healthy individuals (intraocular pressure <22 mm Hg, healthy optic disc by stereophotograph assessment, standard visual fields within normal limits) and 90 eyes of 45 clinic patients (ocular hypertensive, glaucomatous optic neuropathy by stereophotograph assessment and/or repeatable abnormal visual fields) enrolled in the University of California, San Diego Diagnostic Innovations in Glaucoma Study (DIGS) were evaluated. Average mean deviation of patient eyes on standard automated perimetry was −1.81 dB (SD=2.61). Methods The PERG was recorded using the PERGLA paradigm from both eyes simultaneously twice (ie, 2 trials) by a single operator with electrodes being removed and reattached between recordings. Repeatability of PERG amplitude (μV) and phase (π rad) between 2 runs within a single trial (within-trial condition) was compared with repeatability between 2 trials (ie, after electrode replacement, between-trial condition) by calculating the coefficients of variability (CVs) and the intraclass correlation coefficients (ICCs) and displaying Bland-Altman plots. Results For healthy eyes, amplitude CVs (SD) were 11.5% (11.5) and 9.9% (0.79) for within-trial and between-trial conditions, respectively. ICCs were 0.91 and 0.85. Phase CVs were 1.3% (1.5) (within-trials) and 1.5% (1.4) (between-trials) and ICCs were 0.85 and 0.88. For patient eyes, amplitude CVs (SD) were 12.2% (10.1) and 11.2% (7.5) for within-trial and between-trial conditions, respectively. ICCs were 0.92 and 0.89. Phase CVs were 2.2% (2.2) (within-trials) and 2.4% (2.2) (between-trials) and ICCs were 0.82 and 0.83. Bland-Altman plots indicated good agreement between the repeated recordings and were similar within-trials and between-trials for healthy and patient eyes. Conclusions Repeatability of PERGLA recordings is good and is similar within-trials and between-trials for both healthy and patient eyes suggesting this technique is promising for monitoring change over time.Keywords
This publication has 19 references indexed in Scilit:
- Reproducibility of Pattern Electroretinogram in Glaucoma Patients with a Range of Severity of Disease with the New Glaucoma ParadigmOphthalmology, 2008
- ISCEV standard for clinical pattern electroretinography—2007 updateDocumenta Ophthalmologica, 2007
- A New Pattern Electroretinogram Paradigm Evaluated in Terms of User Friendliness and Agreement with PerimetryOphthalmology, 2007
- Habituation of Retinal Ganglion Cell Activity in Response to Steady State Pattern Visual Stimuli in Normal SubjectsInvestigative Ophthalmology & Visual Science, 2005
- Pattern electroretinogram abnormality and glaucomaOphthalmology, 2005
- Normative data for a user-friendly paradigm for pattern electroretinogram recordingOphthalmology, 2004
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- The Pattern-Elicited ElectroretinogramOphthalmologica, 1986
- Pattern ERG in the monkey after section of the optic nerveExperimental Brain Research, 1985
- Electroretinographic Responses to Alternating Gratings Before and After Section of the Optic NerveScience, 1981