Biomechanical Strength of the Peri-Loc® Proximal Tibial Plate: A Comparison of All-Locked Versus Hybrid Locked/Nonlocked Screw Configurations
Authors:
- Chris EstesCEChris Estes
- Peter RheePRPeter Rhee
- M Wade ShraderMSM Wade Shrader
- Kristine CsavinaKCKristine Csavina
- Marc C JacofskyMJMarc C Jacofsky
- David J JacofskyDJDavid J Jacofsky
- 1 May 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 22 (5) , 312-316
Abstract
The purpose of this study was to compare the biomechanical properties of a contoured locking plate instrumented with either an all-locked or hybrid locked/nonlocked screw construct in a proximal metaphyseal fracture of the tibia (AO 41-A3.2). A standardized proximal metaphyseal wedge osteotomy (AO 41-A3.2) was created in five pairs of cadaveric tibia. Each pair was randomly instrumented with either an all-locked or combination locked/nonlocked screw construct using a locked contoured periarticular plate (Peri-Loc periarticular locked plating system, Smith & Nephew, Memphis, TN). Vertical subsidence (irreversible deformation) and deflection (reversible deformation) in each pair were analyzed and compared. Load to failure, defined by complete fracture gap closure, was also determined. There was no statistically significant difference in vertical subsidence (P = 0.19) or deflection (P = 0.19) of the proximal tibia between the all-locked and combination locked/nonlocked screw construct with increasing levels of cyclical axial load from 200 to 1200 N. Failure occurred at a mean value of 2160 N in the locked group and 1760 N in the hybrid group (P = 0.19); the failure mode was plate bending in all specimens. The results indicate that the use of compression screws with locked screws in this particular construct allows a similar amount of irreversible and reversible deformation in response to an axial load when compared to an all-locked screw construct. This suggests that there is no statistically significant difference in the stability in fixation between the two methods, allowing the surgeon the freedom to choose the appropriate screw combination unique to each fracture.Keywords
This publication has 14 references indexed in Scilit:
- Biomechanics of Locked Plates and ScrewsJournal of Orthopaedic Trauma, 2004
- Neue Stabilisierungsverfahren an der proximalen Tibia (LISS/LCP)Therapeutische Umschau, 2003
- Guidelines for the clinical application of the LCPInjury, 2003
- Less invasive stabilization system (LISS) for fractures of the proximal tibia: Indications, Surgical Technique and Preliminary Results of the UMC Clinical TrialInjury, 2003
- Bicondylar Tibial Plateau Fractures: A Biomechanical StudyClinical Orthopaedics and Related Research, 2003
- Extraosseous Blood Supply of the Tibia and the Effects of Different Plating Techniques: A Human Cadaveric StudyJournal of Orthopaedic Trauma, 2002
- Biomechanical Evaluation of the Less Invasive Stabilization System for the Internal Fixation of Distal Femur FracturesJournal of Orthopaedic Trauma, 2001
- Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and boneInjury, 2000
- Minimally Invasive Plate Osteosynthesis: Does Percutaneous Plating Disrupt Femoral Blood Supply Less Than the Traditional Technique?Journal of Orthopaedic Trauma, 1999
- Interfragmentary lag screws and external fixation for severe open tibial fractures: a biomechanical and clinical studyClinical Biomechanics, 1994