Paralyzable and non-paralyzable dead-time corrections for the neutron detectors of the KUCA facility driven by external neutron sources
- 4 September 2019
- journal article
- research article
- Published by Taylor & Francis in Journal of Nuclear Science and Technology
- Vol. 57 (2) , 157-168
Abstract
This study examines dead-time corrections for the neutron detectors of the Kyoto University Critical Assembly (KUCA) driven by a stationary or pulsed external neutron sources. For the stationary (californium) external neutron source, the MCNP simulations producing the detector timestamps and the MATLAB scripts correcting for the dead-time effect have been validated with analytical formulae from literature. For the pulsed external neutron source (produced by a proton accelerator), the experimental detector timestamps have been corrected for the dead-time effect using the aforementioned MATLAB scripts utilizing the backward extrapolation method and the paralyzable or non-paralyzable dead-time model. Then, the keff obtained from the dead-time corrected timestamps has been corrected for the detector position and type. Finally, the corrected experimental keff has been compared with the one obtained from MCNP criticality simulations.Keywords
This publication has 11 references indexed in Scilit:
- Application of the backward extrapolation method to pulsed neutron sourcesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2018
- Advances in the computation of the Sjöstrand, Rossi, and Feynman distributionsProgress in Nuclear Energy, 2017
- Dead time corrections using the backward extrapolation methodNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2017
- Neutron Characteristics of Solid Targets in Accelerator-Driven System with 100-MeV Protons at Kyoto University Critical AssemblyNuclear Technology, 2015
- Perspectives of research and development of accelerator-driven system in Kyoto University Research Reactor InstituteProgress in Nuclear Energy, 2015
- Measurement of dead time by time interval distribution methodNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Monte Carlo and deterministic calculation of the Bell and Glasstone spatial correction factorComputer Physics Communications, 2012
- Impact of the neutron detector choice on Bell and Glasstone spatial correction factor for subcriticality measurementNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012
- JENDL-4.0: A New Library for Nuclear Science and EngineeringJournal of Nuclear Science and Technology, 2011
- Static and Kinetic Experiments on Accelerator-Driven System with 14MeV Neutrons in Kyoto University Critical AssemblyJournal of Nuclear Science and Technology, 2008