Abstract
We describe a parallel pair of high voltage electric field plates designed and constructed to minimise magnetic Johnson noise. They are formed by laminating glass substrates with a commercially available polyimide (Kapton) tape, covered with a thin gold film. Tested in vacuum, the outgassing rate is less than 5 × 10−5 mbar l/s. The plates have been operated at electric fields up to 8.3 kV/cm, when the leakage current is at most a few hundred pA. The design is discussed in the context of a molecular spin precession experiment to measure the permanent electric dipole moment of the electron.
All Related Versions
Funding Information
  • Engineering and Physical Sciences Research Council
  • Science and Technology Facilities Council
  • Royal Society
  • European Research Council