Monte Carlo model of the Studsvik BNCT clinical beam: Description and validation
- 18 November 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (12) , 3107-3117
Abstract
The neutron beam at the Studsvik facility for boron neutron capture therapy (BNCT) and the validation of the related computational model developed for the MCNP‐4B Monte Carlo code are presented. Several measurements performed at the epithermal neutron port used for clinical trials have been made in order to validate the Monte Carlo computational model. The good general agreement between the MCNP calculations and the experimental results has provided an adequate check of the calculation procedure. In particular, at the nominal reactor power of 1 MW, the calculated in‐air epithermal neutron flux in the energy interval between 0.4 eV–10 keV is (±1.2% 1 std. dev.) while the measured value is (±5.0% 1 std. dev.). Furthermore, the calculated in‐phantom thermal neutron flux, equal to (±1.0% 1 std. dev.), and the corresponding measured value of (±5.3% 1 std. dev.) agree within their respective uncertainties. The only statistically significant disagreement is a discrepancy of 39% between the MCNP calculations of the in‐air photon kerma and the corresponding experimental value. Despite this, a quite acceptable overall in‐phantom beam performance was obtained, with a maximum value of the therapeutic ratio (the ratio between the local tumor dose and the maximum healthy tissue dose) equal to 6.7. The described MCNP model of the Studsvik facility has been deemed adequate to evaluate further improvements in the beam design as well as to plan experimental work.Keywords
This publication has 29 references indexed in Scilit:
- Reference dosimetry at the neutron capture therapy facility at StudsvikMedical Physics, 2003
- Controllability of depth dose distribution for neutron capture therapy at the Heavy Water Neutron Irradiation Facility of Kyoto University Research ReactorMedical Physics, 2002
- Photon quality correction factors for ionization chambers in an epithermal neutron beamPhysics in Medicine & Biology, 2002
- Reference dosimetry calculations for neutron capture therapy with comparison of analytical and voxel modelsMedical Physics, 2002
- On the optimal energy of epithermal neutron beams for BNCTPhysics in Medicine & Biology, 1999
- Selective delivery of 10B to soft tissue sarcoma using 10B-L-borophenylalanine for boron neutron capture therapy.British Journal of Radiology, 1998
- Characteristics of Mg/Ar ionisation chambers used as gamma-ray dosemeters in mixed neutron-photon fieldsPhysics in Medicine & Biology, 1986
- Current status of 10B-neutron capture therapy: Enhancement of tumor dose via beam filtration and dose rate, and the effects of these parameters on minimum boron content: A theoretical evaluationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1985
- Photon mass attenuation and energy-absorption coefficientsThe International Journal of Applied Radiation and Isotopes, 1982
- Kerma factors of elements and compounds for neutron energies below 30 MeVThe International Journal of Applied Radiation and Isotopes, 1982