Abstract
Amyloid-β oligomers (AβO) have become a representative biomarkers for early diagnosis of Alzheimer's disease. Here, we report on an aptasensor based on stem-loop probes for a sensitive and specific detection of Aβ oligomers by an am-perometric transducer principle using alternating current voltammetry (ACV). Stem-loop probes with redox-active moieties are immobilized on a gold substrate as receptor element. The signal transduction mechanism relies on redox ferrocene (Fc) reporting via charge transfer on a molecular recognition event involving a conformational change of the molecular beacon. The stem-loop structures were optimized by considering the aptamers’ stem length, spacer, and dif-ferent ferrocene terminals. In addition, the assembly and signal recording including aptamer concentration and ACV frequency are discussed. Using the optimized stem-loop probe (B-3’ Fc), the aptasensor showed an increase of the Fc peak current induced by AβO binding within the broad concentration range spanning six orders of magnitude. Furthermore, the detection limit of the sensor can be further decreased by optimizing the ACV frequency, however at the costs of a narrowed detection range. In this work a label-free electrochemical aptasensor is demonstrated that facilitates the quantification of the concentration of Aβ oligomers with high selectivity and subpicomolar sensitivity, which may be conducive to improve the diagnosis and pharmacology studies of Alzheimer's disease.
Funding Information
  • China Scholarship Council (201506890024)

This publication has 33 references indexed in Scilit: