Performance of antiscatter grids in diagnostic radiology: Experimental measurements and Monte Carlo simulation studies
Authors:
- Heang‐Ping ChanHCHeang‐Ping Chan
- Yoshiharu HigashidaYHYoshiharu Higashida
- Kunio DoiKDKunio Doi
- 1 July 1985
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 12 (4) , 449-454
Abstract
We have devised an experimental method with which one can accurately measure the transmission of primary radiation and the transmission of total radiation by an antiscatter grid in a setting similar to a practical radiographic examination. We measured the transmission values for 27 combinations of x‐ray tube potentials, phantom thicknesses, screen–film systems, and grid parameters. The standard deviation of one measurement was estimated to be 2.9% and 1.5% for the total and primary transmissions, respectively. The measured grid transmission was compared with results predicted by our Monte Carlo calculations; 92% of the measured and calculated values agree within two standard deviations. This close agreement indicates that our Monte Carlo calculation can accurately predict the performance of antiscatter grids under diagnostic imaging conditions.Keywords
Funding Information
- U.S. Public Health Service (CA 24806)
This publication has 4 references indexed in Scilit:
- Ultra-high-strip-density radiographic grids: a new antiscatter technique for mammography.Radiology, 1985
- Physical and clinical evaluation of new high-strip-density radiographic grids.Radiology, 1983
- The validity of Monte Carlo simulation in studies of scattered radiation in diagnostic radiologyPhysics in Medicine & Biology, 1983
- Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100Atomic Data and Nuclear Data Tables, 1970