Potential dosimetric error in the adaptive workflow of a 1.5 T MR-Linac from patient movement relative to immobilisation systems
Open Access
- 16 January 2024
- journal article
- research article
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 47 (1) , 351-359
Abstract
In magnetic resonance- (MR-) based adaptive workflows for an MR-linac, the treatment plan is optimized and recalculated online using the daily MR images. The Unity MR-linac is supplied with a patient positioning device (ppd) using pelvic and abdomen thermoplastic masks attached to a board with high-density components. This study highlights the dosimetric effect of using this in such workflows when there are relative patient-ppd displacements, as these are not visualized on MR imaging and the treatment planning system assumes the patient is fixed relative to the ppd. The online adapted plans of two example rectum cancer patients treated at a Unity MR-linac were perturbed by introducing relative patient-ppd displacements, and the effect was evaluated on plan dosimetry. Forty-eight perturbed clinical adapted plans were recalculated, based on online MR-based synthetic computed tomography, and compared with the original plans, using dose-volume histogram parameters and gamma analysis. The target volume covered by the prescribed dose ( $${\text{D}}_{\text{p}\text{r}\text{e}}$$ ) and by at least 107% of $${\text{D}}_{\text{p}\text{r}\text{e}}$$ varied up to − 1.87% and + 3.67%, respectively for 0.5 cm displacements, and to − 3.18% and + 4.96% for 2 cm displacements; whilst 2%–2 mm gamma analysis showed a median value of 92.9%. The use of a patient positioning system with high-density components in a Unity MR-based online adaptive treatment workflow can introduce unrecognized errors in plan dosimetry and it is recommended not to use such a device for such treatments, without modifying the device and the workflow, followed by careful clinical evaluation, or alternatively to use other immobilization methods.
Keywords
Funding Information
- Natural Science Foundation of Sichuan Province (2023NSFSC0711)
- Medical Engineering Innovation Fund for Cancer (ZYGX2021YGCX002)
This publication has 29 references indexed in Scilit:
- Normal Tissue Constraints for Abdominal and Thoracic Stereotactic Body RadiotherapySeminars in Radiation Oncology, 2017
- Assessment of three-dimensional setup errors in image-guided pelvic radiotherapy for uterine and cervical cancer using kilovoltage cone-beam computed tomography and its effect on planning target volume marginsJournal of Cancer Research and Therapeutics, 2017
- Comparative analysis of thermoplastic masks versus vacuum cushions in stereotactic body radiotherapyRadiation Oncology, 2015
- The Use of Dual Vacuum Stabilization Device to Reduce Kidney Motion for Stereotactic Radiotherapy PlanningTechnology in Cancer Research & Treatment, 2014
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Interfractional Variations in Patient Setup and Anatomic Change Assessed by Daily Computed TomographyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Comparison of daily megavoltage electronic portal imaging or kilovoltage imaging with marker seeds to ultrasound imaging or skin marks for prostate localization and treatment positioning in patients with prostate cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Randomized trial on two types of thermoplastic masks for patient immobilization during radiation therapy for head-and-neck cancerInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumorsRadiotherapy and Oncology, 2001
- Accuracy of pelvic radiotherapy: prospective analysis of 90 patients in a randomised trial of blocked versus standard radiotherapyRadiotherapy and Oncology, 1996