Polymer thick film technology for improved simultaneous dEEG/MRI recording: Safety and MRI data quality
- 15 February 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 77 (2) , 895-903
Abstract
Purpose: To develop a 256‐channel dense‐array electroencephalography (dEEG) sensor net (the Ink‐Net) using high‐resistance polymer thick film (PTF) technology to improve safety and data quality during simultaneous dEEG/MRI.Methods: Heating safety was assessed with temperature measurements in an anthropomorphic head phantom during a 30‐min, induced‐heating scan at 7T. MRI quality assessment used B1 field mapping and functional MRI (fMRI) retinotopic scans in three humans at 3T. Performance of the 256‐channel PTF Ink‐Net was compared with a 256‐channel MR‐conditional copper‐wired electroencephalography (EEG) net and to scans with no sensor net. A visual evoked potential paradigm assessed EEG quality within and outside the 3T scanner.Results: Phantom temperature measurements revealed nonsignificant heating (ISO 10974) in the presence of either EEG net. In human B1 field and fMRI scans, the Ink‐Net showed greatly reduced cross‐modal artifact and less signal degradation than the copper‐wired net, and comparable quality to MRI without sensor net. Cross‐modal ballistocardiogram artifact in the EEG was comparable for both nets.Conclusion: High‐resistance PTF technology can be effectively implemented in a 256‐channel dEEG sensor net for MR conditional use at 7T and with significantly improved structural and fMRI data quality as assessed at 3T. Magn Reson Med 77:895–903, 2017. © 2016 International Society for Magnetic Resonance in MedicineKeywords
Funding Information
- National Institute of Neurological Disorders and Stroke (R44NS071988, U01 NS075026)
- National Institute of Biomedical Imaging and Bioengineering (R21EB016449)
- National Center for Research Resources (P41RR14075)
This publication has 50 references indexed in Scilit:
- Identifying the sources of the pulse artefact in EEG recordings made inside an MR scannerNeuroImage, 2013
- Functional imaging of seizures and epilepsyCurrent Opinion in Neurology, 2012
- Electroencephalographic source imaging: a prospective study of 152 operated epileptic patientsBrain, 2011
- 96‐Channel receive‐only head coil for 3 Tesla: Design optimization and evaluationMagnetic Resonance in Medicine, 2009
- An open-source hardware and software system for acquisition and real-time processing of electrophysiology during high field MRIJournal of Neuroscience Methods, 2008
- Safety of localizing epilepsy monitoring intracranial electroencephalograph electrodes using MRI: Radiofrequency‐induced heatingJournal of Magnetic Resonance Imaging, 2008
- A comparative study of different artefact removal algorithms for EEG signals acquired during functional MRINeuroImage, 2007
- MR heating tests of MR critical implantsJournal of Magnetic Resonance Imaging, 2007
- EEG/(f)MRI measurements at 7 Tesla using a new EEG cap (“InkCap”)NeuroImage, 2006
- The dielectric properties of biological tissues: I. Literature surveyPhysics in Medicine & Biology, 1996