Practical approach to method verification in plasma and validation in cerebrospinal fluid under accreditation using a flexible scope in molecular virology: setting up the HIV, HBV and HCV Aptima™ Quant Dx assays
Authors:
- 24 August 2019
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 58 (2) , 251-260
Abstract
Background: Our laboratory obtained the ISO 15189 accreditation for the plasmatic HIV-1, HBV and HCV viral load (VL) using the m2000 RealTime™ system, which was recently changed for the platform Panther®. Here, we discuss a strategy for performing method validation/verification very quickly.Methods: We performed the mandatory (repeatability, internal quality assessment [IQA], measurement uncertainty [MU]) and optional technical verifications for CE/IVD assays using the flexible scope range A. We also performed the mandatory assays for the validation of HIV-1 VL in the cerebrospinal fluid (CSF) using the flexible scope range B. The change was checked by following up on the turnaround time (TAT).Results: The coefficient of variation (CV%) for repeatability and IQA complied with the limit of 0.25 log. The MU results ranged from 0.04 to 0.25 log copies or IU/mL. The comparisons of methods showed excellent correlations (R2 = 0.96 for the three parameters) but a delayed centrifugation on HCV VL showed variations of up to 2 log IU/mL. An excellent linearity for HIV-1 in the CSF was obtained from 1.5 to 5 log copies/mL with R2 = 0.99. The TAT increased (84%–98%) in routine usage.Conclusions: The three Aptima assays are well suited for routine laboratory use and can be integrated within less than 2 weeks in accordance with flexible scope range A. Our data allows us to confidently perform HIV-1 VL in CSF following flexible scope range B. Finally, we provide an organizational guide for flexible scope management in molecular virology within a short time frame.Keywords
This publication has 28 references indexed in Scilit:
- Performance of the New Aptima HCV Quant Dx Assay in Comparison to the Cobas TaqMan HCV2 Test for Use with the High Pure System in Detection and Quantification of Hepatitis C Virus RNA in Plasma or SerumJournal of Clinical Microbiology, 2016
- Aptima HIV-1 Quant Dx—A fully automated assay for both diagnosis and quantification of HIV-1Journal of Clinical Virology, 2016
- Evaluation of Virological Microparticle Enzyme Immunoassay According to the ISO 15189: Real-Life Experience in a University HospitalClinical Laboratory, 2016
- Accreditation process in European countries – an EFLM surveyClinical Chemistry and Laboratory Medicine (CCLM), 2016
- Flexible scope for ISO 15189 accreditation: a guidance prepared by the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) Working Group Accreditation and ISO/CEN standards (WG-A/ISO)Clinical Chemistry and Laboratory Medicine (CCLM), 2015
- Validation of carbapenemase and extended-spectrum -lactamase multiplex endpoint PCR assays according to ISO 15189Journal of Antimicrobial Chemotherapy, 2013
- Discordance Between Cerebral Spinal Fluid and Plasma HIV Replication in Patients with Neurological Symptoms Who Are Receiving Suppressive Antiretroviral TherapyClinical Infectious Diseases, 2010
- Verification and validation of diagnostic laboratory tests in clinical virologyJournal of Clinical Virology, 2007
- Monitoring Plasma HIV-1 RNA Levels in Addition to CD4+ Lymphocyte Count Improves Assessment of Antiretroviral Therapeutic ResponseAnnals of Internal Medicine, 1997
- HIV viral load markers in clinical practiceNature Medicine, 1996