Systematic review and meta-analysis of side effects and severe complications caused by intracoronary pressure measurement
Open Access
- 17 November 2025
- journal article
- review article
- Published by Springer Nature in Clinical Research in Cardiology
- p. 1-8
Abstract
Background: Even though current guidelines recommend intracoronary pressure measurement such as fractional flow reserve (FFR) to assess the hemodynamic relevance of angiographically intermediate stenoses, little is known about the incidence of side effects and there is no large-scale data on complications of intracoronary pressure measurements. Methods: The PubMed database was searched for large multicenter trials and registries of patients undergoing intracoronary pressure measurement. Publications were screened regarding side effects and complications. Data on chest pain, dyspnea, arrhythmia, bronchospasm, hypotension, nausea, and coronary artery injury were collected and analyzed. Results: A total of 19 studies with 22,768 patients were included. FFR measurement was reported to cause chest pain in 29% (95% CI 2–91%), dyspnea in 20%, hypotension in 0.8% (95% CI 0.3–2.1%), vomiting or nausea in 0.9%, and bronchospasm in 0.2% (95% CI 0.0–0.7%) of patients. Arrhythmia occurred in 3% (95% CI 1–16%), more specifically atrioventricular block in 2% (95% CI 0–14%) and ventricular arrhythmia in 0.1% (95% CI 0.1–0.3%), during intracoronary pressure measurement. The pressure wire was reported to have caused coronary artery dissection in 0.2% (95% CI 0.1–0.5%), coronary artery occlusion in 0.2% (95% CI 0.1–0.3%), and coronary artery perforation in 0.1% of patients. Coronary artery injury required revascularization on rare occasions and was fatal in two patients. Conclusions: Transient side effects due to adenosine administration are common when FFR is measured. While infrequent, coronary artery injury caused by the pressure wire occurs in 0.5% of patients and is a relevant complication of intracoronary pressure measurement. GraphicalKeywords
Funding Information
- State of Bavaria (ER-MIKKI, LSM-2007-0002)
- Universitätsklinikum Erlangen
This publication has 36 references indexed in Scilit:
- Use of coronary physiology to guide revascularization in clinical practice: results of the F(FR)2 registryClinical Research in Cardiology, 2024
- PRISMA2020 : An R package and Shiny app for producing PRISMA 2020‐compliant flow diagrams, with interactivity for optimised digital transparency and Open SynthesisCampbell Systematic Reviews, 2022
- A global registry of fractional flow reserve (FFR)–guided management during routine care: Study design, baseline characteristics and outcomes of invasive managementCatheterization and Cardiovascular Interventions, 2020
- Fractional flow reserve versus angiography guided percutaneous coronary intervention: An updated systematic reviewCatheterization and Cardiovascular Interventions, 2017
- Correlation between Angiographic and Physiologic Evaluation of Coronary Artery Narrowings in Patients With Aortic Valve StenosisThe American Journal of Cardiology, 2017
- Safety of guidewire-based measurement of fractional flow reserve and the index of microvascular resistance using intravenous adenosine in patients with acute or recent myocardial infarctionInternational Journal of Cardiology, 2015
- The use of the acute Pd/Pa drop after intracoronary nitroglycerin infusion to rule out significant FFR: CANICA (Can intracoronary nitroglycerin predict fractional flow reserve without adenosine?) multicenter studyCatheterization and Cardiovascular Interventions, 2015
- Prevalence of visual–functional mismatch regarding coronary artery stenosis in the CVIT-DEFER registryCardiovascular Intervention and Therapeutics, 2014
- Diagnostic Accuracy of Fractional Flow Reserve From Anatomic CT AngiographyJAMA, 2012
- Development and Validation of a New Adenosine-Independent Index of Stenosis Severity From Coronary Wave–Intensity AnalysisJACC, 2012