Measuring Normal Stress Differences in Polymer Melt Shear Flow

Authors:

Abstract
This paper describes modifications to a commercial cone‐and‐plate shear rheometer (Rheometrics Mechanical Spectrometer RMS 800) that allow the measurement of the two normal stress differences during shear flow of polymer melts at temperatures presently up to 250°C. As a first step, a new heating system has been developed that maintains the sample at a constant temperature within less than 0.01°C variance. This reduces the fluctuations of the normal force signal, caused by thermal expansion of the polymer melt between the cone and the plate, down to the fluctuations at room temperature. Using this heater, the measurement of the transient normal force (and therefore the first normal stress difference) is as easy as the measurement of the transient shear stress. Additionally, the temperature rise from room temperature to stable conditions at test temperature takes only 30 minutes, and requires only a small amount of pure nitrogen or argon to provide an inert environment for the sample. In the second step, the plate has been partitioned into an inner disk connected to the torque/normal force transducer and an outer ring fixed to the frame of the instrument. With this configuration the transducermeasures only the normal force portion F i , acting on the inner disk of radius R i , whereas the total normal force exerted by the sample of radius R>R i is F. The ratio F i /F reflects the ratio of the two normal stress differences N 2 /N 1 . By testing samples of different radii R at the same shear rate, the two normal stress differences including their time response can be separated. The tests were performed with a low‐density polyethylene (LDPE) at 150°C. At the low shear rate of 0.5 s −1 , the ratio |N 2 /N 1 | is 0.24, and the two normal stress differences have opposite signs.
Keywords