Value of a cinefluoroscopic assessment of the Lillehei-Kaster prosthetic heart values
Authors:
- 1 October 1978
- journal article
- research article
- Published by Springer Nature in CardioVascular and Interventional Radiology
- Vol. 1 (4) , 229-232
Abstract
In order to assess the functioning of the pivoting disc of the Lillehei-Kaster prosthetic valve, a cinefluoroscopic technique using multiple angulations was developed. The disc was detected in 240 (99%) of the 242 valves studied (126 aortic, 107 mitral, and nine tricuspid valves). The maximal-opening angle of the disc was measured in 75% of the aortic but only 20% of the mitral valve prostheses in the 182 valve studies in which this was attempted. Prosthetic dysfucntions were suspected clinically in three aortic and three tricuspid valves and confirmed with the cinefluoroscopic technique. This cinefluoroscopic approach constitutes a rapid, noninvasive, and sensitive evalaution of the function of the Lillehei-Kaster prosthetic valve in the aortic position.Keywords
This publication has 8 references indexed in Scilit:
- Phonocardiographic and echocardiographic features of Lillehei-Kaster mitral prosthesisJournal of Clinical Ultrasound, 1977
- Clinical experience with the Lillehei-Kaster cardiac valve prosthesisThe Journal of Thoracic and Cardiovascular Surgery, 1976
- In Vivo Evaluation of the Lillehei-Kaster Heart Valve ProsthesisThe Annals of Thoracic Surgery, 1976
- Heart-valve replacement with Lillehei-Kaster pivoting disk prosthesis.1974
- [Lillehei-Kaster heart valve prosthesis: short-term evaluation in 25 patients].1974
- Echocardiographic and Phonocardiographic Characteristics of the Lillehei-Kaster Mitral Valve ProsthesisCirculation, 1974
- A new central flow pivoting disc aortic and mitral prosthesis: Initial clinical experienceThe American Journal of Cardiology, 1970
- The Lillehei-Kaster pivoting disc aortic prosthesis and a comparative study of its pulsatile flow characteristics with four other prostheses.1970