The Medoff sliding plate and a standard sliding hip screw for unstable intertrochanteric fractures: A mechanical comparison in cadaver femurs
Authors:
Open Access
- 1 January 1998
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 69 (3) , 266-272
Abstract
The Medoff sliding plate has a dual side capability along both the femoral shaft and neck to increase theoretically interfragmentary compression and load-sharing in hip fractures. We studied intertrochanteric fracture fixation in cadaveric bone to determine whether this device has a mechanical advantage over a standard sliding hip screw. 2-part and 4-part fractures were created in 12 cadaver femurs. The fractures were fixated and sequentially destabilized; bone and plate strains and fragment displacements were determined during testing, as a function of applied physiological loads before and after short-term cycling. The Medoff sliding plate imposed a higher mean medial cortex strain than the sliding hip screw in all fracture models and at all loading levels, and the difference was statistically significant in the 2-part and in the unstable 4-part fracture models. The loading of the medial cortex region after cycling was approximately 50% higher in the Medoff samples than in the sliding hip screw samples. There were no significant differences in plate strains, fracture displacements or load to failure between the 2 devices. These observations favor the dual sliding principle as regards providing fracture compression and load-sharing, which may explain low failure rates in clinical series of unstable intertrochanteric fractures, treated with the Medoff sliding plate.Keywords
This publication has 19 references indexed in Scilit:
- How to minimize failures of fixation of unstable intertrochanteric fracturesInjury, 1995
- A Biomechanical Evaluation of the Long Stem Intramedullary Hip ScrewJournal of Orthopaedic Trauma, 1995
- One-way compression along the femoral shaft with the Medoff sliding plate:The first European experience of 104 intertrochanteric fractures with a 1-year follow-upActa Orthopaedica, 1995
- Functional outcome after intertrochanteric fractures of the femur: does the implant matter? A prospective study of 100 consecutive casesInjury, 1993
- A Biomechanical Analysis of the Sliding Hip ScrewJournal of Orthopaedic Trauma, 1990
- Sliding hip screws: modes of failureInjury, 1989
- Fixation of unstable trochanteric hip fractures: A cadaver study comparing three different devicesActa Orthopaedica, 1988
- Unstable Trochanteric Fractures Treated with the Sliding Screw-Plate System:A Biomechanical Study of Unstable Trochanteric FracturesIIIActa Orthopaedica, 1978
- TREATMENT OF INTERTROCHANTERIC HIP FRACTURES WITH A COMPRESSION HIP SCREW AND A NAIL PLATE1976
- Trochanteric Femoral Fractures Treated with McLaughlin OsteosynthesisActa Orthopaedica, 1975