Mid-IR laser ablation of articular and fibro-cartilage:A wavelength dependence study of thermal injury and crater morphology
Open Access
- 1 March 2006
- journal article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 38 (3) , 218-228
Abstract
Background and Objective The aim of this study was to evaluate areas of collateral thermal injury and crater morphology for evidence of wavelength‐dependent effects on the ablation of articular cartilage and fibro‐cartilage (meniscus) using selected mid‐IR wavelengths produced by a free electron laser. Study Design/Materials and Methods Two types of cartilage, articular cartilage and fibro‐cartilage were used in the study. The wavelengths (λ) evaluated were 2.79, 2.9, 6.1, and 6.45 µm generated by a free electron laser (FEL) using a 4 microseconds macropulse configuration. The zone of thermal injury and crater morphology produced by laser ablation were examined by light microscopy following standard histologic processing. Results The zone of thermal injury and crater morphology created in cartilage by the FEL at selected mid‐IR wavelengths were examined as a function of incident radiant exposure. Ablation using λ = 6.1 µm provided the largest crater size for both articular and fibro‐cartilage at all radiant exposures. For the zones of collateral thermal injury in articular cartilage, λ = 6.1 µm produced the least thermal injury at the radiant exposure of 7.6 J/cm2. When the radiant exposure is increased to 20.4 J/cm2, both λ = 6.1 and 6.45 µm produced less thermal injury than the ablation using λ = 2.79 and 2.9 µm. The greatest amount of collateral thermal injury was produced by λ = 2.79 µm for both tissue types. Conclusions The results demonstrate that crater depth and collateral thermal injury produced in articular cartilage and fibro‐cartilage are wavelength‐dependent with 6.1 µm providing the largest craters at all radiant exposures. The least amount of thermal injury was created in articular cartilage using λ = 6.1 µm at the radiant exposure of 7.6 J/cm2. Both 6.1 and 6.45 µm wavelengths demonstrated similar amount of thermal injury at 20 J/cm2 that was less than λ = 2.79 and 2.9 µm at similar fluences. These observations are explained based on the absorption by water and protein in the tissue types studied. It is further observed that the use of crater dimensions may not provide a reliable estimate for the amount of tissue removal provided by an ablation procedure. Lasers Surg. Med. 38:218–228, 2006.Keywords
This publication has 29 references indexed in Scilit:
- The effect of free-electron laser pulse structure on mid-infrared soft-tissue ablation: biological effectsPhysics in Medicine & Biology, 2005
- Free electron laser ablation of articular and fibro-cartilage at 2.79, 2.9, 6.1, and 6.45 ?m: Mass removal studiesLasers in Surgery and Medicine, 2005
- Free-electron-laser-based biophysical and biomedical instrumentationReview of Scientific Instruments, 2003
- Mechanisms of Pulsed Laser Ablation of Biological TissuesChemical Reviews, 2003
- Quantification and modeling of the dynamic changes in the absorption coefficient of water at λ = 2.94 μmIEEE Journal of Selected Topics in Quantum Electronics, 2001
- Pulsed Ho:YAG laser meniscectomy: Effect of pulsewidth on tissue penetration rate and lateral thermal damageLasers in Surgery and Medicine, 1995
- 308 nm Excimer laser ablation of cartilageLasers in Surgery and Medicine, 1994
- Thermal side effects of fiber‐guided XeCl excimer laser drilling of cartilageLasers in Surgery and Medicine, 1994
- Comparison of the erbium‐yttrium aluminum garnet and carbon dioxide lasers for in vitro bone and cartilage ablationThe Laryngoscope, 1990
- CartilageScientific American, 1984