Performance of three different continuous glucose monitoring systems in children with type 1 diabetes during a diabetes summer camp
Open Access
- 21 November 2020
- journal article
- research article
- Published by Wiley in Pediatric Diabetes
- Vol. 22 (2) , 271-278
Abstract
Aim The aim of this study was to assess accuracy of the three most commonly used continuous glucose monitoring (CGM) systems in almost real‐life situation during a diabetes camp in children with type 1 diabetes (T1D) aged 9‐14 years. Methods Data was gathered during a 2‐week summer camp under physicians’ supervision. Out of 38 participating children with T1D (aged: 11.0 [9.9; 12.1] years; 57% girls, mean HbA1c 7.2 [6.9; 7.7] %,) 37 wore a CGM system (either Abbott FreeStyle Libre (FSL), Dexcom G6 (DEX) or Medtronic Enlite (ENL)) throughout the camp. All concomitantly available data pairs of capillary glucose measurements and sensor values were used for the analysis. Mean absolute relative difference (MARD) was calculated and Parkes Error Grid analyses were done for all three systems used. Results In total 2,079 data pairs were available for analysis. The overall MARDs of CGM systems used at the camp was FSL: 13.3% (6.7;21.6). DEX: 10.3% (5.8; 16.7) and ENL 8.5% (3.6; 15.6). During eu‐, hypo‐ and hyperglycemia MARDs were lowest in ENL. Highest MARDs were seen in hypoglycemia, where all three systems exhibited MARDs above 15%. Overnight MARDs of all systems was higher than during day‐time. Conclusion All sensors performed worst in hypoglycemia. Performance of the adequately calibrated Medtronic system outperformed the factory‐calibrated sensors. For clinical practice, it is important to adequately train children with T1D and families in the correct procedures for sensors that require calibrations.Keywords
This publication has 35 references indexed in Scilit:
- Comparative Accuracy of 17 Point-of-Care Glucose MetersJournal of Diabetes Science and Technology, 2016
- Continuous Glucose Monitoring in Patients With Type 1 Diabetes Using Insulin InjectionsDiabetes Care, 2016
- Accuracy and user performance evaluation of the Contour® Next Link 2.4 blood glucose monitoring systemClinica Chimica Acta, 2015
- A Novel Method to Detect Pressure-Induced Sensor Attenuations (PISA) in an Artificial PancreasJournal of Diabetes Science and Technology, 2014
- Accuracy of two continuous glucose monitoring systems: a head‐to‐head comparison under clinical research centre and daily life conditionsDiabetes, Obesity and Metabolism, 2014
- Lot-to-Lot Variability of Test Strips and Accuracy Assessment of Systems for Self-Monitoring of Blood Glucose according to ISO 15197Journal of Diabetes Science and Technology, 2012
- Efficacy of Continuous Glucose Monitoring in Improving Glycemic Control and Reducing Hypoglycemia: A Systematic Review and Meta-Analysis of Randomized TrialsJournal of Diabetes Science and Technology, 2011
- Development of a WHO growth reference for school-aged children and adolescents2007
- A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.Diabetes Care, 2000
- Smoothing reference centile curves: The lms method and penalized likelihoodStatistics in Medicine, 1992