Signal Deconvolution Concept Combined With Cubic Spline Interpolation to Improve Timing With Phoswich PET Detectors
Authors:
- 16 June 2009
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 56 (3) , 581-587
Abstract
PET imaging scanners based on all-digital architecture offer greater data processing flexibility and the possibility to recalibrate the system with simple software procedures. The LabPET scanner, designed at the Universite de Sherbrooke, is one such device. It is built around dual LYSO/LGSO scintillators in a phoswich arrangement coupled to Avalanche Photodiodes (APD) and combined with highly parallel readout and processing electronics. This approach enables the implementation of advanced real time digital signal processing methods to compute energy resolution, crystal identification and the arrival time of events. Timing extraction represents the highest challenge because of the low sampling frequency (45 MHz), the quantization error and the presence of a Zero Order Hold (ZOH) in the system. The aim of this paper is to present a method to increase coincidence time accuracy by adequately addressing each of these limitations. The proposed method uses a Deconvolution concept based on adaptive filter theory preceded by Cubic Spline interpolation to improve digital timing performance. The method achieves 4.5 ns, 8.2 ns, and 6.5 ns timing resolution with LYSO-LYSO, LGSO-LGSO and mixed crystals phoswich coincidences, respectively.Keywords
This publication has 16 references indexed in Scilit:
- Real Time Implementation of a Wiener Filter Based Crystal Identification AlgorithmIEEE Transactions on Nuclear Science, 2008
- Wavelets-Based Crystal Identification of Phoswich Detectors for Small-Animal PETIEEE Transactions on Nuclear Science, 2008
- Timing Improvement by Low-Pass Filtering and Linear Interpolation for the LabPET ScannerIEEE Transactions on Nuclear Science, 2008
- The data acquisition system of ClearPET neuro - a small animal PET scannerIEEE Transactions on Nuclear Science, 2006
- High-resolution digital online linear procedure for timing of detected eventsIEEE Transactions on Nuclear Science, 2006
- Time discrimination techniques using artificial neural networks for positron emission tomography2005
- Pulse shape discrimination of LSO and LuYAP scintillators for depth of interaction detection in PETIEEE Transactions on Nuclear Science, 2003
- Inter-crystal scatter in a dual layer, high resolution LSO-APD positron emission tomographPhysics in Medicine & Biology, 2003
- Towards real-time digital pulse processing based on least-mean-squares algorithmsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997
- Integrated constant-fraction discriminator shaping techniques for the PHENIX lead-scintillator calorimeterIEEE Transactions on Nuclear Science, 1997