Biomechanical Considerations in Plate Osteosynthesis: The Effect of Plate-to-Bone Compression With and Without Angular Screw Stability
Authors:
- Karl StoffelKSKarl Stoffel
- Kai-Uwe LorenzKLKai-Uwe Lorenz
- Markus S KusterMKMarkus S Kuster
- 1 July 2007
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 21 (6) , 362-368
Abstract
We compared the biomechanical stability of bone-plate constructs using a compression plate (CP), an internal fixator (IF), and a combination plate (CP/IF). Standardized simulated shaft fractures with a segmental defect in composite bones (n=60) and intraarticular distal femur fractures with a comminuted supracondylar zone in fresh frozen cadaveric femurs (n=36) were stabilized by CP, IF, and CP/IF. Construct stiffness, plastic deformation, and fixation strength were measured under axial compression and torsion using a biaxial testing machine. The experimental results indicate for the distal femur fracture model that IF has less loss of reduction by plastic deformation under axial load compared to CP (IF 61% of CP). Under torsion, the CP showed significantly (P<0.05) decreased plastic deformation compared to the IF (CP 51% of IF). The combination (CP/IF) of the 2 fixation principles generally resulted in a higher load to failure under axial compression and torsion (145% failure load of CP and 118% of IF under axial compression, 88% of CP and 109% of IF under torsion). Results were similar between the 2 fracture models. Under compression, IF provides similar fixation in comminuted fractures and was better than the CP for avoiding loss of reduction, whereas under torsional loading, CP was more important for stiffness, plastic deformation, and load to failure than IF. However, combination (CP/IF) fixation seems advisable in intraarticular and extraarticular fractures of long bones with a metaphyseal comminution. These data may be utilized by surgeons to build a more specific treatment plan in patients with these fracture types.Keywords
This publication has 27 references indexed in Scilit:
- The Mechanical Behavior of Locking Compression Plates Compared With Dynamic Compression Plates in a Cadaver Radius ModelJournal of Orthopaedic Trauma, 2005
- Assessment of Stiffness and Strength of 4 Different Implants Available for Equine Fracture Treatment: A Study on a 20° Oblique Long‐Bone Fracture Model Using a Bone SubstituteVeterinary Surgery, 2005
- Delayed and non-union of the humeral diaphysis—compression plate or internal plate fixator?Injury, 2004
- A PROSPECTIVE, RANDOMIZED TRIAL COMPARING THE LIMITED CONTACT DYNAMIC COMPRESSION PLATE WITH THE POINT CONTACT FIXATOR FOR FOREARM FRACTURESJournal of Bone and Joint Surgery, 2003
- Development of the Locking Compression PlateInjury, 2003
- The LCP-concept in the operative treatment of distal humerus fractures – biological, biomechanical and surgical aspectsInjury, 2003
- Biomechanical testing of the LCP – how can stability in locked internal fixators be controlled?Injury, 2003
- LISS PLT: Design, mechanical and biomechanical characteristicsInjury, 2003
- Biomechanical Evaluation of the Less Invasive Stabilization System for the Internal Fixation of Distal Femur FracturesJournal of Orthopaedic Trauma, 2001
- The concept of biological plating using the limited contact — dynamic compression plate (LC-DCP)Injury, 1991