Proton beam energy calibration of the 3 MV Tandetron™ at IFIN-HH
Authors:
- D.A. TestovDTD.A. Testov
- M. StraticiucMSM. Straticiuc
- P.-A. SöderströmPSP.-A. Söderström
- S. AogakiSAS. Aogaki
- R.-F. AndreiRAR.-F. Andrei
- A.K. AzhibekovAAA.K. Azhibekov
- D.L. BalabanskiDBD.L. Balabanski
- S-R. BanSBS-R. Ban
- M. BrezeanuMBM. Brezeanu
- I. BurduceaIBI. Burducea
- P. ConstantinPCP. Constantin
- A. DhalADA. Dhal
Open Access
- 1 July 2025
- journal article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 20 (07) , P07051
Abstract
The ELIGANT-TN long neutron counter, the ELISSA silicon strip array and the Mini-eTPC active gas target detector all represent nuclear physics research tools constructed at Extreme Light Infrastructure - Nuclear Physics (ELI-NP) for use in γ-beam experiments. Currently, these three setups are used at the charged particle accelerators of the “Horia Hulubei” National Institute for R&D in Physics and Nuclear Engineering (IFIN-HH), for performing reaction cross-section measurements. Accurate knowledge of the incident beam energy is of paramount importance for such studies, which is why in the present article we report on a new proton beam energy calibration performed using the following threshold reactions: 7Li(p,n)7Be, 13C(p,n)13N and 51V(p,n)51Cr. The measurements were performed by varying the projectile energy across the key range of ∼ 1600–3200 keV. The results demonstrate a linear dependence of the projectile energy on the terminal voltage.Keywords
This publication has 8 references indexed in Scilit:
- Cross section measurements of low-energy charged particle induced reactions using moderated neutron counter arraysEPJ Web of Conferences, 2023
- ELIGANT-GN — ELI Gamma Above Neutron Threshold: The Gamma-Neutron setupNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2022
- The energy calibration of the IFIN-HH 3 MV Tandetron accelerator for alpha particlesAIP Advances, 2021
- Energy calibration of ion-beam accelerators with neutron and gamma nuclear reactions between 2 MeV and 4 MeVNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2021
- Performance studies of X3 silicon detectors for the future ELISSA array at ELI-NPJournal of Instrumentation, 2018
- Nuclear Astrophysics at ELI-NP: the ELISSA prototype tested at Laboratori Nazionali del SudEPJ Web of Conferences, 2017
- A new ion beam facility based on a 3 MV Tandetron™ at IFIN-HH, RomaniaNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2015
- Thresholds for () Reactions on 26 Intermediate-Weight NucleiPhysical Review B, 1964