Applying a Deep Learning Network in Continuous Physiological Parameter Estimation Based on Photoplethysmography Sensor Signals
- 8 November 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Sensors Journal
- Vol. 22 (1) , 385-392
Abstract
In this paper, we propose a continuous physiological parameter estimation model based on a deep learning network for photoplethysmography (PPG) sensor signals. Signals of 8-s duration were incorporated into the proposed model in this study for frequent estimation of the systolic blood pressure (BP), diastolic BP, heart rate (HR), and mean arterial pressure of the human body; this facilitated early identification and monitoring of physiological conditions and thus reduced the risk of cardiovascular disease. The proposed model was designed using a convolutional neural network (CNN) and long short-term memory (LSTM) network. This model was trained and validated using the large-scale Multiparameter Intelligent Monitoring in Intensive Care database. The CNN was used to extract features from PPG signals automatically. This automatic extraction replaced the conventional manual feature extraction process. Features with time-series were then analyzed using the LSTM network to estimate physiological parameters. Subsequently, ten-fold cross-validation was conducted to reveal the mean absolute errors ± standard deviations of participants’ systolic BP, diastolic BP, HR, and mean arterial pressure to be 2.54 ± 3.88, 1.59 ± 2.45, 1.62 ± 2.55, and 1.59 ± 2.34 mmHg, respectively. These values meet the standards established by the Association for the Advancement of Medical Instrumentation and the British Hypertension Society. The proposed method facilitates the accurate, continuous monitoring of the BP and HR.Keywords
Funding Information
- Ministry of Science and Technology (MOST 108-2221-E-150-022-MY3)
- National Taiwan Ocean University
This publication has 41 references indexed in Scilit:
- A Robust Heart Rate Monitoring Scheme Using Photoplethysmographic Signals Corrupted by Intense Motion ArtifactsIEEE Transactions on Biomedical Engineering, 2015
- Cuff-less high-accuracy calibration-free blood pressure estimation using pulse transit time2015
- Optimizing mean arterial pressure in septic shock: a critical reappraisal of the literatureCritical Care, 2015
- Photoplethysmography-Based Heart Rate Monitoring in Physical Activities via Joint Sparse Spectrum ReconstructionIEEE Transactions on Biomedical Engineering, 2015
- TROIKA: A General Framework for Heart Rate Monitoring Using Wrist-Type Photoplethysmographic Signals During Intensive Physical ExerciseIEEE Transactions on Biomedical Engineering, 2014
- Hypertension in developing countriesThe Lancet, 2012
- Ambulatory Blood-Pressure MonitoringNew England Journal of Medicine, 2006
- Age-specific relevance of usual blood pressure to vascular mortality: a meta-analysis of individual data for one million adults in 61 prospective studiesThe Lancet, 2002
- PhysioBank, PhysioToolkit, and PhysioNetCirculation, 2000
- Robust Estimation of a Location ParameterThe Annals of Mathematical Statistics, 1964