Design and Evaluation of a Catheter Contact-Force Controller for Cardiac Ablation Therapy
- 4 February 2016
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 63 (11) , 2301-2307
Abstract
Goal: Maintaining a constant contact force (CF) of an ablation catheter during cardiac catheter ablation therapy is clinically challenging due to inherent myocardial motion, often resulting in poor ablation of arrhythmogenic substrates. To enable a prescribed contact force to be applied during ablation, a catheter contact force controller (CCFC) was developed. Methods: The system includes a hand-held device attached to a commercial catheter and steerable sheath. A compact linear motor assembly attaches to an ablation catheter and autonomously controls its relative position within the shaft of the steerable sheath. A closed-loop control system is implemented within embedded electronics to enable real-time catheter-tissue contact force control. To evaluate the performance of the CCFC, a linear motion phantom was used to impose a series of physiological CF profiles; lesion CF was controlled at prescribed levels ranging from 15 to 40 g. Results: For a prescribed CF of 25 g, the CCFC was able to regulate the CF with a root mean squared error of 3.7 ± 0.7 g. The ability of the CCFC to retract the catheter upon sudden changes in tissue motion, which may have caused tissue damage, was also demonstrated. Finally, the device was able to regulate the CF for a predetermined amount of time according to a force-time integral model. Conclusion: The developed CCFC is capable of regulating catheter-tissue CF in a laboratory setting that mimics clinical ablation therapy. Significance: Catheter-tissue CF control promises to improve the precision and success of ablation lesion delivery.Keywords
Funding Information
- Canadian Institutes of Health Research (MOP-115184)
- Ontario Research Fund
- NSERC Collaborative Research and Training Experience Program in Computer-Assisted Medical Interventions
- University of Western Ontario
- Career Investigator
- Heart and Stroke Foundation of Ontario
This publication has 20 references indexed in Scilit:
- Use of contact force sensing technology during radiofrequency ablation reduces recurrence of atrial fibrillation: A systematic review and meta-analysisHeart Rhythm, 2015
- Characterization of Catheter–Tissue Contact Force During Epicardial Radiofrequency Ablation in an Ovine ModelCirculation: Arrhythmia and Electrophysiology, 2013
- A Magnetic-Resonance-Imaging-Compatible Remote Catheter Navigation SystemIEEE Transactions on Biomedical Engineering, 2013
- Efforts to enhance catheter stability improve atrial fibrillation ablation outcomeHeart Rhythm, 2013
- Effect of respiration on catheter-tissue contact force during ablation of atrial arrhythmiasHeart Rhythm, 2012
- Force control of flexible catheter robots for beating heart surgery2011
- Steerable Versus Nonsteerable Sheath Technology in Atrial Fibrillation AblationCirculation: Arrhythmia and Electrophysiology, 2011
- General anesthesia reduces the prevalence of pulmonary vein reconnection during repeat ablation when compared with conscious sedation: Results from a randomized studyHeart Rhythm, 2011
- Novel Contact Force Sensor Incorporated in Irrigated Radiofrequency Ablation Catheter Predicts Lesion Size and Incidence of Steam Pop and ThrombusCirculation: Arrhythmia and Electrophysiology, 2008
- Steerable Sheath Catheter Navigation for Ablation of Atrial Fibrillation: A Case‐Control StudyPacing and Clinical Electrophysiology, 2008