Rapid scanning serves to mitigate the problems of scintillation and dynamic range for Fourier spectrometry, as well as to simplify the measurement procedures. Rapid scanning means scanning at a rate such that the interferometric fringes act as their own chopper. It also means scanning at a rate such that the fringe frequencies exceed the scintillation frequencies, thereby avoiding confusion between the two. Of necessity the scanning becomes repetitive for the measurement of faint sources. During observation the repetitive interferogram signals are recorded on one channel of a stereo magnetic tape recorder, while timing information is recorded on the other channel. The interferograms are subsequently digitized and summed (in synchronism according to the timing channel) into a magnetic core memory. Fourier analysis of the resulting interferogram can be finally carried out by either analogue techniques or by digital computation. Astronomical spectra obtained by two instruments, one in the visible (S-20) range, the other in the lead sulphide range, are shown to demonstrate the efficacy of rapid scanning. Finally a curious systematic effect of erroneous digitizing to enhance weak spectral absorption lines was discovered and will be illustrated