Aerosol Sampling: Comparison of Two Rotating Impactors for Field Droplet Sizing and Volumetric Measurements
Authors:
- Mike J. GreerMGMike J. Greer
- Bradley K. FritzBFBradley K. Fritz
- W. Clint HoffmannWHW. Clint Hoffmann
- 1 December 2009
- journal article
- research article
- Published by American Mosquito Control Association in Journal of the American Mosquito Control Association
- Vol. 25 (4) , 474-479
Abstract
This article compares the collection characteristics of a new rotating impactor Florida Latham Bonds (FLB) sampler for ultrafine aerosols with a mimic of the industry standard (Hock-type). The volume and droplet-size distribution collected by the rotating impactors were measured via spectroscopy and microscopy. The rotary impactors were colocated with an isokinetic air sampler for a total volume flux measurement and a laser diffraction instrument for droplet-size distribution measurement. The measured volumetric flux and droplet-size distribution collection efficiencies were compared across 3 wind speeds (1, 1.8, and 3.5 m/sec). The FLB sampler had higher flux collection efficiencies than the Hock-type sampler. The FLB sampler collected 89%, 87%, and 98% of the total volume available per unit area at 1, 1.8, and 3.5 m/sec, respectively, whereas the Hock-type sampler collected 68%, 19%, and 21% of across the same wind speeds. Changes in wind speed had less impact and resulted in less data variability for the FLB sampler.Keywords
This publication has 4 references indexed in Scilit:
- Canopy Effects Droplet Size Distribution and Meteorological ChangeJournal of the American Mosquito Control Association, 2008
- Development of a System for Determining Collection Efficiency of Spray SamplersApplied Engineering in Agriculture, 2008
- An evaluation of two field samplers for monitoring spray driftCrop Protection, 1996
- The Impaction of Aerosol Particles on Cylinders, Spheres, Ribbons and DiscsAnnals of Occupational Hygiene, 1967