The Difference in the Assessment of Knee Extension/Flexion Angles during Gait between Two Calibration Methods for Wearable Goniometer Sensors
Open Access
Abstract
Frontal and axial knee motion can affect the accuracy of the knee extension/flexion motion measurement using a wearable goniometer. The purpose of this study was to test the hypothesis that calibrating the goniometer on an individual’s body would reduce errors in knee flexion angle during gait, compared to bench calibration. Ten young adults (23.2 ± 1.3 years) were enrolled. Knee flexion angles during gait were simultaneously assessed using a wearable goniometer sensor and an optical three-dimensional motion analysis system, and the absolute error (AE) between the two methods was calculated. The mean AE across a gait cycle was 2.4° (0.5°) for the on-body calibration, and the AE was acceptable (<5°) throughout a gait cycle (range: 1.5–3.8°). The mean AE for the on-bench calibration was 4.9° (3.4°) (range: 1.9–13.6°). Statistical parametric mapping (SPM) analysis revealed that the AE of the on-body calibration was significantly smaller than that of the on-bench calibration during 67–82% of the gait cycle. The results indicated that the on-body calibration of a goniometer sensor had acceptable and better validity compared to the on-bench calibration, especially for the swing phase of gait.Keywords
Funding Information
- Nitto Denko Corporation, Tokyo, Japan (PC87200005)
This publication has 43 references indexed in Scilit:
- Epidemiology of locomotive syndrome using updated clinical decision limits: 6-year follow-ups of the ROAD studyJournal of Bone and Mineral Metabolism, 2022
- Wearable pendant device monitoring using new wavelet-based methods shows daily life and laboratory gaits are differentMedical & Biological Engineering & Computing, 2015
- Quantified self and human movement: A review on the clinical impact of wearable sensing and feedback for gait analysis and interventionGait & Posture, 2014
- Biomechanical Deviations During Level Walking Associated With Knee Osteoarthritis: A Systematic Review and Meta‐AnalysisArthritis Care & Research, 2013
- A systematic review of the effect of ageing and falls history on minimum foot clearance characteristics during level walkingGait & Posture, 2010
- A new formulation of the coefficient of multiple correlation to assess the similarity of waveforms measured synchronously by different motion analysis protocolsGait & Posture, 2010
- First in vivo assessment of “Outwalk”: a novel protocol for clinical gait analysis based on inertial and magnetic sensorsMedical & Biological Engineering & Computing, 2009
- The reliability of three-dimensional kinematic gait measurements: A systematic reviewGait & Posture, 2009
- The Functional Flexion-Extension Axis of the Knee Corresponds to the Surgical Epicondylar AxisThe Journal of Arthroplasty, 2005
- Prediction of hip joint centre location from external landmarksHuman Movement Science, 1989