Do the tube current and metal artifact reduction influence the diagnosis of vertical root fracture in a tooth positioned in the vicinity of a zirconium implant? A CBCT study
- 22 August 2020
- journal article
- research article
- Published by Springer Nature in Clinical Oral Investigations
- Vol. 25 (4) , 2229-2235
Abstract
Aim To evaluate the influence of the tube current and metal artifact reduction (MAR) tool on the diagnosis of vertical root fractures (VRF) in a tooth adjacent to a zirconium implant, in cone-beam computed tomography (CBCT) images. Methodology Thirty single-rooted teeth (15 with VRF and 15 control group) were individually positioned in a mandible, and scanned with the OP300 CBCT unit. Images were acquired using a standardized protocol: 5 × 5 cm field of view, 0.08-mm voxel size, and 90 kVp. Each tooth was scanned with and without a zirconium implant in its vicinity, using different tube currents (4 mA, 8 mA, and 10 mA) and conditions of MAR (enabled × disabled). Diagnostic values were calculated for each protocol, and compared by multi-way analysis of variance. Results The ROC curve and sensitivity values did not differ significantly among the tube currents, regardless of the presence of the implant and MAR condition (p > 0.05). There were also no significant differences among the tube currents for the specificity values (p > 0.05); however, the specificity differed significantly between the “with implant” and “without implant” conditions, within the same MAR condition and tube current (p < 0.05). Specificity was significantly lower when the implant was present (p < 0.05). Conclusion The presence of a zirconium implant impairs the diagnosis of VRF in teeth adjacent to the artifact-generator material. Neither the tube current nor the MAR tool is effective in improving this diagnostic task. Therefore, in this clinical scenario, the use of the lowest tube current (4 mA), without MAR activation, is recommended. Clinical relevance Considering that the tube current is one of the main factors that influence the radiation dose and image quality in CBCT, and that metal artifacts negatively influence the diagnosis of VRF in areas adjacent to the artifact-generator material, it is important to evaluate the effect of this energetic parameter in the diagnosis of VRF in teeth adjacent to zirconium implants.Keywords
Funding Information
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (001)
This publication has 27 references indexed in Scilit:
- Cone-beam Computed Tomography for Detecting Vertical Root Fractures in Endodontically Treated Teeth: A Systematic ReviewJournal of Endodontics, 2016
- Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal postsDentomaxillofacial Radiology, 2015
- Influence of CBCT enhancement filters on diagnosis of vertical root fractures: a simulation study in endodontically treated teeth with and without intracanal postsDentomaxillofacial Radiology, 2015
- Comparing the Effect of Different Voxel Resolutions for Assessment of Vertical Root-Fracture of Permanent TeethIranian Journal of Radiology, 2015
- Prevalence of vertical root fracture as the reason for tooth extraction in dental clinicsClinical Oral Investigations, 2014
- Evaluation of Cone-beam Computed Tomography in the Diagnosis of Vertical Root Fractures: The Influence of Imaging Modes and Root Canal MaterialsJournal of Endodontics, 2014
- A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation doseDentomaxillofacial Radiology, 2014
- Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction modeDentomaxillofacial Radiology, 2013
- Cone beam CT scans with and without artefact reduction in root fracture detection of endodontically treated teethDentomaxillofacial Radiology, 2013
- Metal artefact reduction with cone beam CT: anin vitrostudyDentomaxillofacial Radiology, 2012