Analysis and Realization of Adaptive Dual-Comb Spectroscopy

Authors:
  • 杨力杨力YangLi
    杨力YangLi
  • 沈旭沈旭玲ShenXuling
    沈旭玲ShenXuling
  • 杨康杨康文YangKangwen
    杨康文YangKangwen
  • 郝强郝强HaoQiang
    郝强HaoQiang
  • 曾和曾和平ZengHeping
    曾和平ZengHeping

Abstract
In this paper, partial P branch lines in the v, absorption band of the acetylene arc measured by the adaptive dual-comb spectroscopy sampling technique. The validity of the method is verified by comparison of the measured results with the standard absorption lines in HITRAN database. The mechanism of adaptive sampling technique based on the sampling theorem is analyzed. High-order features and bandwidth characteristics of the adaptive sampling techniques arc further discussed. In the experiment setup, 200 nm super continuums arc generated in nonlinear fiber, and then the wide spectra arc extended. Partial absorption lines of the acetylene arc retrieved from a 600 is pulse signal in time domain with 1.09 GHz spectral resolution at a refresh rate of 180 Hz. The fast instabilities between combs arc well compensated by the adaptive sampling technique under a room temperature environment. Meanwhile, with a 7.01 MHz stabilization accuracy of repetition rates, we suppress slow drifts effectively by further stabilizing the repetition rates of combs and tuning the center frequency of continuouswave lasers. In this way, the stable running time of the dual-comb adaptive sampling system is increased from only a few minutes to more than half an hour, and thus the long-term stability and practicability of adaptive dual-comb spectroscopy arc enhanced.