A comparative assessment of contamination rates in gastrointestinal endoscope reprocessing: sterilization versus high‐level disinfection
Open Access
Abstract
Objectives: This study aimed to evaluate the disparity in culture results between sterilization and high‐level disinfection (HLD) for duodenoscopes and linear endoscopic ultrasound (EUS), and to assess the effectiveness of different bacterial contamination detection methods.Methods: This is a prospective randomized study, including duodenoscopes and linear EUS with adenosine triphosphate bioluminescence assay values below 200 relative light units after manual cleaning which were randomly assigned to undergo either sterilization or HLD in a 1:1 ratio. Following disinfection, all endoscopes were subjected to adenosine triphosphate bioluminescence assay testing and cultures using both swab and liquid samples from endoscope channels.Results: Totally 752 endoscopes (444 duodenoscopes and 308 linear EUS) were studied. After disinfection, the positive culture rates for the sterilization and HLD groups were 5.9% and 7.2%, respectively (p = 0.460). No significant difference in contamination rates was observed between duodenoscopes and linear EUS (5.9% and 7.5%, respectively; p = 0.379), and no significant association between contamination rates and the presence of biliary stones was seen (7.3% vs. 6.9%; p = 0.613). The detection rate of bacteria from liquid samples taken from endoscope channels was 0.5%, which was significantly lower than the swabbing method (6.0%, p < 0.001).Conclusions: This study found no statistically significant difference in contamination rates between sterilization and HLD methods for gastrointestinal endoscope reprocessing. The type of endoscope and the presence of biliary stones did not influence the positive culture rate. The swabbing method showed significantly higher bacterial detection when compared with liquid samples.Keywords
This publication has 25 references indexed in Scilit:
- Disinfection and Sterilization in Health Care FacilitiesInfectious Disease Clinics of North America, 2016
- Impact of ethylene oxide gas sterilization of duodenoscopes after a carbapenem-resistant Enterobacteriaceae outbreakGastrointestinal Endoscopy, 2016
- Practical toolkit for monitoring endoscope reprocessing effectiveness: Identification of viable bacteria on gastroscopes, colonoscopes, and bronchoscopesAmerican Journal of Infection Control, 2016
- Disinfection and sterilization: An overviewAmerican Journal of Infection Control, 2013
- Transmission of Infection by Flexible Gastrointestinal Endoscopy and BronchoscopyClinical Microbiology Reviews, 2013
- Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channelsAmerican Journal of Infection Control, 2013
- The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channelsAmerican Journal of Infection Control, 2013
- Evaluation of ATP Bioluminescence Assays for Potential Use in a Hospital SettingInfection Control & Hospital Epidemiology, 2011
- ESGE–ESGENA guideline: Cleaning and disinfection in gastrointestinal endoscopyEndoscopy, 2008
- ESGE-ESGENA guideline for quality assurance in reprocessing: Microbiological surveillance testing in endoscopyEndoscopy, 2007