Surface Damage Versus Tibial Polyethylene Insert Conformity: A Retrieval Study
- 1 July 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 470 (7) , 1814-1825
Abstract
Surface damage of the tibial polyethylene insert in TKA is thought to diminish with increasing conformity, based on computed lower contact stresses. Added constraint from higher conformity may, however, result in greater forces in vivo. We therefore determined whether increased conformity was associated with increased surface pitting, delamination, creep, and polishing in a group of retrieved tibial inserts. We compared 38 inserts with a dished articular surface (conforming group) with 31 inserts that were unconstrained and nonconforming in the sagittal plane (less conforming group). The two groups had identical polyethylene composition and processing history. The articulating surfaces were scored for pitting, delamination, deformation/creep, and polishing. Evidence of edge loading and the presence of embedded bone cement were also recorded. The conforming inserts were associated with higher delamination and pitting scores but lower polishing scores, even after adjusting for the effects of sex, age, insert thickness, and implantation duration. Long implantation duration and male sex were also associated with increased delamination, pitting, and polishing, whereas long shelf life was associated only with increased delamination. The conforming group also had approximately a fourfold greater prevalence of edge loading and approximately a threefold greater prevalence of embedded bone cement. The latter was associated with higher scores and proportions of delamination and pitting. These findings suggest more conformity may increase surface fatigue damage in TKA. Higher constraint-induced stresses during secondary motions and more possibility for edge loading and bone cement capture on a dished surface may account for these results. The selection of materials with high fatigue resistance may be particularly important for high-conformity/constraint tibial inserts. In addition, awareness of the benefits and trade-offs with conformity may allow better matching of TKA design to patient.Keywords
This publication has 48 references indexed in Scilit:
- In Vivo Oxidation Contributes to Delamination but not Pitting in Polyethylene Components for Total Knee ArthroplastyThe Journal of Arthroplasty, 2011
- What is the Evidence for Total Knee Arthroplasty in Young Patients?: A Systematic Review of the LiteratureClinical Orthopaedics and Related Research, 2011
- Increased Conformity Offers Diminishing Returns for Reducing Total Knee Replacement WearJournal of Biomechanical Engineering, 2010
- Knee model of hydrodynamic lubrication during the gait cycle and the influence of prosthetic joint conformityJournal of Orthopaedic Science, 2009
- Tibial Post Wear in Posterior-stabilized Knee Replacements is Design-dependentClinical Orthopaedics and Related Research, 2008
- The combined role of wear particles, macrophages and lymphocytes in the loosening of total joint prosthesesJournal of The Royal Society Interface, 2008
- Why Are Total Knee Arthroplasties Failing Today?Clinical Orthopaedics and Related Research, 2002
- Relationship between body mass index and activity in hip or knee arthroplasty patientsJournal of Orthopaedic Research, 2000
- Tractive forces during rolling motion of the knee: Implications for wear in total knee replacementJournal of Biomechanics, 1997
- Polyethylene wear in unicondylar knee prosthesesActa Orthopaedica, 1992