Broadband measurement system for the specific absorption rate above 300 MHz
Authors:
- 1 August 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 88-90
Abstract
The specific absorption rate (SAR) is an indicator of the thermal effects caused by microwave exposure in biological tissues. Electric field probe scanning has been used as a practical SAR evaluation technology until now. A measurement standard is necessary for any physical quantity. In general, the existence of a standard constructed with a physically different procedure is important in uncertainty evaluation of the measurement standard. In this report, a thermal SAR measurement method using an optical fiber thermal sensor is proposed. The thermal SAR measurement method is not suitable for dosimetric assessments for an actual cellular phone. Nevertheless, a thermal sensor works independently of the frequency, and is therefore suitable for primary testing of an electric field probe scanning system. The results show that the proposed method is able to measure SAR within an error range of 1-13% in frequency range of 300-800 MHz and 2-6 GHz.Keywords
This publication has 2 references indexed in Scilit:
- A Specific Absorption Rate Measurement Method Using Fiber Optic Thermal SensorsIEEE Transactions on Instrumentation and Measurement, 2010
- Energy absorption mechanism by biological bodies in the near field of dipole antennas above 300 MHzIEEE Transactions on Vehicular Technology, 1992