Medium resolution off-axis electron holography with millivolt sensitivity
Authors:
- 1 October 2007
- journal article
- conference paper
- Published by AIP Publishing in Applied Physics Letters
- Vol. 91 (14) , 143501
Abstract
Focused ion beam prepared silicon calibration specimens with boron doped layers have been examined using off-axis electron holography. By using a state-of-the-art FEI Titan microscope with unprecedented stability, we have been able to record holograms for time periods of with contrast levels of almost 40% and an average signal on the charge coupled device camera of 30 000 counts. A consequence of this is a significant improvement of the signal-to-noise ratio in the phase images allowing steps in potential of less than to be measured if sufficient care is taken during specimen preparation.
Keywords
This publication has 6 references indexed in Scilit:
- The influence of electron irradiation on electron holography of focused ion beam milled GaAs p-n junctionsJournal of Applied Physics, 2007
- Improvement in electron holographic phase images of focused-ion-beam-milled GaAs and Si p-n junctions by in situ annealingApplied Physics Letters, 2006
- Quantitative Electron Holography of Biased Semiconductor DevicesPhysical Review Letters, 2002
- Two-Dimensional Mapping of the Electrostatic Potential in Transistors by Electron HolographyPhysical Review Letters, 1999
- Experimental study of amplitude and phase detection limits in electron holographyUltramicroscopy, 1996
- Detection limits in quantitative off-axis electron holographyUltramicroscopy, 1993