Clinical validation of a mandibular movement signal based system for the diagnosis of pediatric sleep apnea

Authors:

Abstract
Background: Given the high prevalence and risk for outcomes associated with pediatric obstructive sleep apnea (OSA), there is a need for simplified diagnostic approaches. A prospective study in 140 children undergoing in‐laboratory polysomnography (PSG) evaluates the accuracy of a recently developed system (Sunrise) to estimate respiratory efforts by monitoring sleep mandibular movements (MM) for the diagnosis of OSA (Sunrise™).Methods: Diagnosis and severity were defined by an obstructive apnea/hypopnea index (OAHI) ≥ 1 (mild), ≥ 5 (moderate), and ≥ 10 events/h (severe). Agreement between PSG and Sunrise™ was assessed by Bland–Altman method comparing respiratory disturbances hourly index (RDI) (obstructive apneas, hypopneas, and respiratory effort‐related arousals) during PSG (PSG_RDI), and Sunrise RDI (Sr_RDI). Performance of Sr_RDI was determined via ROC curves evaluating the device sensitivity and specificity at PSG_OAHI ≥ 1, 5, and 15 events/h.Results: A median difference of 1.57 events/h, 95% confidence interval: −2.49 to 8.11 was found between Sr_RDI and PSG_RDI. Areas under the ROC curves of Sr_RDI were 0.75 (interquartile range [IQR]: 0.72–0.78), 0.90 (IQR: 0.86–0.92) and 0.95 (IQR: 0.90–0.99) for detecting children with PSG_OAHI ≥ 1, PSG_OAHI ≥ 5, or PSG_ OAHI ≥ 10, respectively.Conclusion: MM automated analysis shows significant promise to diagnose moderate‐to‐severe pediatric OSA.
Funding Information
  • Agence Nationale de la Recherche (ANR‐12‐TECS‐0010, ANR‐15‐IDEX‐02, ANR‐19‐P3IA‐0003)
  • National Institutes of Health (AG061824, HL130984, HL140548)

This publication has 41 references indexed in Scilit: